SPIKE FLOW Endocardial Lead Safety and Performance Study

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorFIAB SpA

About this trial

The objective of this study is to assess the safety and performance of the SPIKE FLOW endocardial lead.

Primary objectives are to assess:

* safety in terms of freedom from Serious Adverse Events(SADEs); * performance in terms of Mean Pacing Capture Threshold.

Secondary objectives are to assess:

* safety in terms of freedom from Adverse Events (ADEs); * performance in terms of Mean Pacing Impedance; * performance in terms of Sensed R-wave Amplitude. The study is a prospective, non-randomized, single-arm, multi-center investigation in 131 adult subjects with 3 investigational sites in Italy.

The study is expected to take approximately 1 year to enroll. Enrolled subjects will have visit timepoints at: baseline, lead insertion, lead removal, discharge and 30 days post lead removal.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Subject undergoing a surgical or an interventional procedure, requiring temporary cardiac pacing per investigator's medical judgment or subjects with bradycardia who are under evaluation to determine if the cause is transient or permanent;

Subject is at least 18 years of age;

Subject provides written informed consent to the study.

Disqualifiers

Subject has a co-existing temporary pacing or an implanted active permanent pacemaker or a cardiac defibrillator;

Subject in cardiogenic shock;

Subject is currently in atrial fibrillation;

Subject is pregnant or breastfeeding;

Trial design

Design model

Single group

Treatments tested in this trial

  • Lead insertion

    Procedure/Surgery

    Implantation of the device should be performed as per the manufacturer's Instructions for Use (IFU). Procedural details will be captured as appropriate on the procedural worksheets. Any Adverse Events or Device Deficiencies observed shall also be recorded. The following tests and examinations must be performed immediately after lead insertion: * Physical examination and vital signs (includes Pulse, Blood Pressure); * Laboratory testing per site standard practice; * Electrocardiogram (ECG); * Transthoracic Echocardiogram (TTE); * Adverse events and device deficiencies assessment.

  • lead removal

    Procedure/Surgery

    Removal of the device should be performed as per the manufacturer's IFU. Procedural details will be captured as appropriate on the procedural worksheets. Any adverse events or device deficiencies observed shall also be recorded. The following tests and examinations must be performed immediately after lead removal: * Physical examination and vital signs (includes Pulse, Blood Pressure); * Laboratory testing per site standard practice; * ECG; * TTE; * Adverse event and device deficiencies assessment

Treatment groups

131 Participants
are divided into 1 treatment group
Group A: experimental single-armExperimental treatment 2 interventions

Trial outcomes

Primary outcomes

1

Safety: Freedom from Severe ADEs

Time frame
from insertion up to 30 days post-lead removal
2

Performance: Mean Pacing Capture Threshold (>2V)

Time frame
from insertion to Lead removal (Up to 7 days)

Secondary outcomes

1

Safety: Freedom from ADEs

Time frame
from insertion up to 30 days post-lead removal
2

Performance: Mean Pacing Impedance (Ohms)

Impedance measured while pacing the heart (in multiple postures)

Time frame
from insertion to Lead removal (Up to 7 days)
3

Performance: Sensed R-wave Amplitude

Summary statistics for sensed R-wave amplitude for Lead from insertion through removal

Time frame
from insertion to Lead removal (Up to 7 days)