Accuracy of Moni-Patch Compared With Tympanic Temperature Monitoring in Adults at Risk of Fever

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorMurata Vios, Inc.

About this trial

This single-center, prospective clinical accuracy study evaluates the agreement between the Moni-Patch Temperature Monitoring Patch System and continuous tympanic temperature monitoring in adults at increased risk of developing fever during ICU care. Moni-Patch and an FDA-cleared continuous tympanic temperature sensor will be used concurrently for up to 24 hours. Agreement between the two temperature measurements will be evaluated using Bland-Altman analysis during a predefined febrile measurement window.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

None

Disqualifiers

None

Trial design

Design model

Single group

Treatments tested in this trial

  • Moni-Patch Temperature Monitoring Patch System

    Device

    A non-invasive wearable temperature monitoring sensor applied to the neck over the carotid region. Temperature will be monitored continuously for up to 24 hours.

  • Continuous Tympanic Temperature Monitoring System

    Device

    An FDA-cleared continuous tympanic temperature monitoring system used as the clinical reference temperature measurement. The sensor will be placed in the ear and used concurrently with Moni-Patch for up to 24 hours.

Treatment groups

35 Participants
are divided into 1 treatment group
Group A: Moni-Patch and Tympanic Temperature MonitoringExperimental treatment 2 interventions

Trial outcomes

Primary outcomes

1

Agreement Between Moni-Patch and Tympanic Core Temperature Measurements During Fever

Agreement between Moni-Patch and continuous tympanic temperature measurements will be assessed using Bland-Altman analysis. The analysis will estimate the mean temperature difference (bias) and 95% limits of agreement (LOA). The predefined performance targets are bias within ±0.4°C and limits of agreement within -1.0°C to +1.0°C.

Time frame
During up to 24 hours of continuous monitoring after sensor application; the primary analysis will use the earliest qualifying 30-pair febrile measurement window after the initial 30-minute stabilization period.

Secondary outcomes

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Murata Vios, Inc.

Lead sponsor

University of Minnesota

Collaborator

This trial is not recruiting at the moment. You can still explore other options: