This is an observational, proof-of-concept, feasibility study where 30 preterm infants on bubble CPAP with gestational age \< 32+0 weeks will be recruited from the neonatal intensive care unit (NICU) at the Montreal Children's Hospital.
The study's main goals are:
1. To determine the relationship between ambient bubbling sounds and delivered pressures in preterm infants on bCPAP. 2. To determine the relationship between transmitted bubbling sounds and airway pressures transmitted to the lungs of preterm infants on bCPAP. 3. To develop models to predict delivered and transmitted bCPAP pressures from the acoustic properties of bubbling sounds.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Infants on bCPAP with gestational age < 32+0 weeks
Postmenstrual age between 28+0 and 36+6 weeks at the time of the study
Postnatal age greater than 168 hours (7 days) at the time of the study
On the bubble CPAP device with binasal prongs at the time of the study
Disqualifiers
Infants with known major congenital anomalies
Infants with known congenital heart disorders
Infants with known neuromuscular disease
Infants receiving ventilator-derived CPAP at the time of the study
Trial population
Preterm infants admitted to the neonatal intensive care unit at the Montreal Children's Hospital.
Trial design
Cohort
Prospective
Treatments tested in this trial
External pressure transducer
DeviceThe delivered CPAP pressure will be measured using an ultra-thin, multi-use catheter pressure transducer inserted into a port in the expiratory limb of the bubble CPAP circuit.
Standard microphone
DeviceThe bubble sound of the water tank will be collected with a standard condenser microphone directly affixed to the pole holding the water tank, with a secure clip.
Wireless acoustic sensor
DeviceThe wireless acoustic sensor contains a dual microphone capable of capturing target sounds as well as ambient noise. The frequencies associated with ambient noise will be subtracted to maximize the signal-to-noise ratio of the bubble sound waveform. The wireless sensor will be placed on the suprasternal notch of the infant for monitoring the bubble sounds transmitted to the lungs and secured using a silicone-based tape approved for use in neonates. Data will be transmitted in real-time to a research-dedicated tablet using the Bluetooth Communication Controller and stored for future analysis.
Internal pressure transducer
DeviceThe transmitted CPAP pressure will be measured using an ultra-thin, single-use catheter pressure transducer inserted through the mouth to the level of the infant's nasopharynx. The data will be acquired with a sampling rate of 10kHz and stored for later analysis.
Treatment groups
Trial outcomes
Primary outcomes
Pressure
The mean pressure and the standard deviation of the pressure will be computed for each segment.
External bubble CPAP sounds
Two metrics will be computed: 1. The root mean square (RMS) 2. The power contained between a pre-determined range within each segment; We will determine the range that contains 80% of the signal power, in order to minimize other noises from bubble CPAP signal.
Internal bubble CPAP sounds
1. We will apply biomedical signal processing methods to separate the bubbling sounds from breathing sounds. Then, RMS and power will be computed. 2. We will use the Pearson correlation coefficient to compute the relationship between bubbling sound and pressure metrics. We will evaluate the linear regression models to identify the combination of sound metrics and covariates with the highest predictive accuracy.