Advanced Diffusion Magnetic reSonance Prostate Imaging for Reducing Extraneous Biopsies (ASPIRE) - Calibration Component (Multi-centre Scanner Calibration) and Validation Component (Multi-centre Validation)

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexMale
Age18+
SponsorUniversity College, London

About this trial

In 2019 the UK National Institute for Health and Care Excellence (NICE) updated its guidelines to recommend Magnetic Resonance Imaging (MRI) for all patients prior to biopsy as it permits the targeted sampling of suspicious lesions rather than blind systematic biopsy, thereby improving detection of clinically significant prostate cancer while avoiding some unnecessary biopsies.

However, there is still the major clinical problem of patients who have an 'MRI suspicious' lesion ultimately having no clinically significant cancer on biopsy. It was calculated that even by using MRI, up to 50% of prostate biopsies may be unnecessary. Prostate biopsies carry many risks and burdens for patients so reducing unnecessary biopsies is therefore a critical unmet need in prostate cancer diagnostics.

The ASPIRE study has been designed to enable the safe implementation of advanced diffusion MRI into the prostate cancer diagnostic pathway.

This study will compare 3 different advanced diffusion MRI sequences, Vascular, Extracellular, and Restricted Diffusion for Cytometry in Tumours (VERDICT), fast-VERDICT or Restriction Spectrum Imaging (RSI) to discover which one is superior in identifying patients who can safely avoid a biopsy.

The study is also separated into 3 distinct components. Firstly, there is an optional component which, once these advanced diffusion sequences have been programmed onto MRI scanners, will recruit local volunteers who were already planning to have to routine MRI as a quality assurance measure.

Component 1 (Calibration) is concerned with calibrating the same participants scans over more than one site to enable direct comparison of measurements. Patients under active follow up or previously diagnosed cancer should be enrolled in this component.

Component 2 (Validation) sees a participant receiving a standard MRI scan as well as the new advanced diffusion sequences to evaluate the new tests against the reference standard without altering the standard-of-care management.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Patients aged ≥18 scheduled for a prostate mpMRI (Optional Component)

Willing and able to provide written informed consent (Optional Component, Component 1 and Component 2)

Patients aged ≥18 with a prior mpMRI demonstrating at least one definite lesion suspicious for prostate cancer (Likert or PI-RADS score 4 or 5) and either under active surveillance for previously diagnosed low-risk prostate cancer or awaiting definitive treatment (Component 1)

Willing and able to undergo multiple MRI scans at different hospitals (Component 1)

Disqualifiers

Contraindications to MRI (Optional Component, Component 1 and Component 2)

Inability to provide informed consent (Optional Component, Component 1 and Component 2)

Prostate biopsy within the last 3 months (Component 1) OR prior prostate cancer diagnosis or treatment (Component 2)

Inability to travel to partner hospital site (Component 1)

Trial design

Design model

Parallel

Treatments tested in this trial

  • MRI

    Procedure/Surgery

    The combination of the 3 additional sequences as a part of the research scan will distinguish this intervention from other studies.

Treatment groups

660 Participants
are divided into 3 treatment groups
Group A: Optional Scanner Set-Up ComponentOther 1 intervention
Group B: Component 1 (Calibration)Other 1 intervention
Group C: Component 2 (Validation)Active comparator 1 intervention

Trial outcomes

Primary outcomes

1

Calibration Component Analysis: The goal is to characterise differences in quantitative MRI metrics between scanners and derive calibration adjustments. For each pair of scanners (e.g., Philips vs Siemens), a scatter plot of VERDICT fIC values (or other

Calibration Component Analysis: The goal is to characterise differences in quantitative MRI metrics between scanners and derive calibration adjustments. For each pair of scanners (e.g., Philips vs Siemens), a scatter plot of VERDICT fIC values (or other metric) will be examined across all subjects who were scanned on both. We will calculate the mean difference and 95% limits of agreement (Bland-Altman analysis) to quantify bias and variability. A linear regression will be performed, and the regression equation (slope, intercept) will be used as the calibration formula. If needed, higher-order terms or non-linear fit will be considered, but simplicity is preferred. We will also assess the repeatability of measurements on the same scanner (some participants may have had repeat scans on the same machine or two Philips sites, etc.). The outcome of Calibration Component will be reported descriptively. No formal hypothesis test is applicable, but we will note if the inter-scanner difference

Time frame
5 years

Secondary outcomes

1

Direct comparison between VERDICT, fast VERDICT, and RSI

For each patient, we will have three outcomes (for each advanced method combined with MRI). We will compare, for example, MRI+VERDICT vs MRI+RSI specificity by looking at the subset of patients who have discordant results (one test positive, the other negative, and no cancer or cancer accordingly) using McNemar's test. We will do this pairwise for all combinations (VERDICT vs RSI, VERDICT vs fast VERDICT, RSI vs fast VERDICT). Additionally, we may compare AUC (area under the ROC curve) for the continuous metrics from each technique. The ROC curves will be generated for VERDICT fIC, RSI index, etc., against the outcome, possibly stratified by site or vendor. An averaged (pooled data) ROC comparison will indicate overall performance differences

Time frame
5 years
2

Comparison of calibrated vs uncalibrated threshold performance

This will involve looking at the sensitivity and specificity of, for example, a uniform fIC threshold of 0.41 across all sites vs using each site's calibrated equivalent threshold (e.g. 0.35 on Siemens, etc.). We will report whether calibration improved the alignment of results. A paired comparison of these two approaches in the same dataset will be conducted (each patient can be classified by both methods). We hypothesize calibration will modestly improve specificity or consistency across sites, but this needs confirmation.

Time frame
5 years
3

Subgroup analyses

We will examine performance by site and by scanner vendor. For each site individually, we will replicate the primary analysis to see if VERDICT added benefit at that site (these are the "individually powered" validations - each site with \~150 patients should be powered \~80% to see a similar effect size). This will tell us if any site deviates significantly. We will also aggregate data by vendor (e.g., combine both Siemens sites' data) to see if, say, all Siemens scanners behave similarly vs others. If needed, statistical models like logistic regression with site or vendor as random effects could be used to formally test for heterogeneity, but with a minimum of four sites it may be simpler to descriptively report these results.

Time frame
5 years
4

Negative MRI cohort

Although patients with a negative MRI likely won't have biopsy verification, we will track any who had follow-up MRIs or biopsies. We will calculate the observed outcome at 1 year: how many were found to have cancer later. This is not comparative, but it gives an estimate of false negatives for both standard and advanced MRI (since an advanced MRI might conceivably pick up something in an MRI-negative case; we will see if any such case occurred and what their follow-up showed).

Time frame
5 years

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

University College, London

Lead sponsor

University College London Hospitals

Collaborator

Royal Free Hospital NHS Foundation Trust

Collaborator

Cambridge University Hospitals NHS Foundation Trust

Collaborator

University Hospital Southampton NHS Foundation Trust

Collaborator

Imperial College Healthcare NHS Trust

Collaborator