About this trial
This study aims to evaluate whether an artificial intelligence (AI) based software system can assist surgeons in identifying key anatomical structures during minimally invasive upper urinary tract surgery.
During these procedures, surgeons operate while viewing images from laparoscopic endoscope. Key anatomical structures, such as the kidney, renal artery and vein, ureter, renal pelvis, inferior vena cava, and left reproductive vein, as well as surgical targets such as tumors, may sometimes be difficult to distinguish because of bleeding, surgical smoke, tissue movement, changes in tissue appearance, or obstruction by surgical instruments. The study team has developed an AI-based image recognition system that analyzes surgical video in real time and identifies and outlines relevant anatomical structures on the screen during surgery.
The study has two main goals (primary endpoints):
1. To evaluate how accurately the AI system can identify and delineate key anatomical structures in surgical videos collected from different hospitals, surgeons, surgical procedures, and imaging systems. 2. To evaluate the real-time performance and technical reliability of the AI system in the operating room, including processing latency, frame rate, system failure rate, and the proportion of time during surgery in which AI-generated results are successfully displayed.
The study will include adults aged 18 years or older who undergo one of five types of minimally invasive upper urinary tract surgery: partial nephrectomy, radical nephrectomy, radical nephroureterectomy, inferior vena cava tumor thrombectomy, or pyeloplasty.
The study will be conducted at 11 hospitals in China. In the first phase, surgical videos recorded between 2015 and 2025 will be used to develop, validate, and test the AI system. In the second phase, approximately 100 patients will be prospectively enrolled from internal centers, and the AI-generated anatomical recognition results will be displayed on a monitor during surgery. Overall, approximately 1,600 retrospective surgical cases and 100 prospective cases are planned for inclusion.
The AI system is designed solely as an assistive tool for anatomical recognition. It does not make clinical decisions, provide autonomous surgical instructions, or control any surgical equipment. Surgeons remain fully responsible for all clinical and surgical decisions, and the AI display can be turned off at any time.
All participants will receive standard clinical care. Participation in the study will not involve any additional medications, surgical procedures, or blood tests. Identifying personal information will be removed from the surgical videos and clinical data used for the study, and all study data will be stored securely in accordance with applicable data protection requirements.
As secondary objectives, the study will compare the performance of the AI system with that of surgeons with different levels of surgical experience using a set of challenging surgical images or video frames. The study will also evaluate surgeons' experience with the system, including its usability, clarity of display, and whether the AI-generated information interferes with or affects the surgical workflow.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Age ≥18 years.
Undergoing one of the eligible minimally invasive renal or upper urinary tract surgical procedures, including partial nephrectomy, radical nephrectomy, radical nephroureterectomy, inferior vena cava tumor thrombectomy, or pyeloplasty.
Availability of an intraoperative surgical video suitable for study analysis. For retrospective cases, an existing surgical video recording must be available; for prospectively enrolled cases, the surgical video must be capable of being captured during the procedure.
Availability of sufficient routinely collected perioperative clinical data to characterize the case and support the planned study analyses.
Disqualifiers
Severe video incompleteness, missing key surgical phases, or excessively low resolution that precludes valid analysis;
Explicit refusal by the patient to allow the use of their surgical video or clinical data for research purposes.
For the prospective phase, emergency or critically ill patients for whom participation in the study procedures would be inappropriate or could interfere with timely clinical care.
Any other condition that, in the investigator's judgment, makes the participant or case unsuitable for inclusion in the study.
Trial population
Adults aged 18 years or older who undergo minimally invasive kidney surgery at one of 11 participating hospitals in China, covering five procedures: partial nephrectomy, radical nephrectomy, radical nephroureterectomy, inferior vena cava tumor thrombectomy with radical nephrectomy, and pyeloplasty. The population includes patients operated between January 2015 and May 2025 whose intraoperative videos and perioperative records are analyzed retrospectively (approximately 1,600 planned cases), and patients enrolled consecutively during the prospective real-time validation phase (approximately 100 cases). Eligible patients have intraoperative video suitable for anatomical identification and perioperative clinical data sufficient to record baseline characteristics and study outcomes. Patients with emergency critical conditions, severely incomplete or low-resolution video, or refusal of data use are excluded.
Trial design
Cohort
Other
Treatments tested in this trial
AI-based intra-operative anatomical perception system
DeviceAn AI-based software system that analyzes the live surgical video feed during minimally invasive upper urinary tract surgery and performs real-time identification and delineation of key anatomical structures and surgical targets, including the kidney, renal artery and vein, ureter, renal pelvis, inferior vena cava, tumors, and adjacent organs. The AI-generated results are overlaid on a surgical monitor to provide intraoperative visual assistance to the surgeon. The system functions solely as an assistive visualization tool; it does not make clinical decisions, provide autonomous surgical guidance, or control any surgical equipment. All surgical decisions remain under the responsibility of the operating surgeon, and the AI display can be turned off at any time.
Treatment groups
Trial outcomes
Primary outcomes
Dice coefficient
The Dice similarity coefficient will quantify the overlap between the AI-generated segmentation mask and the expert-reviewed reference mask for each target structure in surgical video frames. Dice = 2 x overlap area/(predicted area + reference area), with areas measured as pixel counts. Values range from 0 to 1: 0 indicates no overlap and 1 indicates perfect overlap. Higher values indicate greater agreement with the reference annotation. Results will be reported overall and for each target structure.
Intersection over Union, IoU
Intersection over Union will quantify the overlap between the AI-generated segmentation mask and the expert-reviewed reference mask for each target structure. IoU = intersection area/union area, where the intersection contains pixels shared by both masks and the union contains pixels belonging to either mask. Values range from 0 to 1: 0 indicates no overlap and 1 indicates perfect overlap. Higher values indicate greater segmentation agreement with the reference annotation. Results will be reported overall and for each target structure.
Precision
Precision will measure the proportion of anatomical structure identifications made by the AI system that are correct according to expert-reviewed reference annotations. Precision = TP/(TP + FP), where TP is the number of correctly identified structures and FP is the number of incorrect or spurious identifications. Values range from 0 to 1. A value of 1 means that all reported positive identifications are correct; lower values indicate a greater proportion of false-positive identifications. Results will be reported overall and for each target structure.
Recall
Recall will measure the proportion of target structures present in the expert-reviewed reference annotations that are correctly identified by the AI system. Recall = TP/(TP + FN), where TP is the number of correctly identified target structures and FN is the number of reference-positive target structures that are missed or incorrectly identified. Values range from 0 to 1. A value of 1 means that all reference-positive targets are correctly identified; lower values indicate a greater proportion of missed targets. Results will be reported overall and for each target structure.
Secondary outcomes
an investigator-developed Likert scale questionnaire
After each evaluated prospective surgical procedure, surgeons will complete the investigator-developed Artificial Intelligence System Usability and Surgeon Experience Questionnaire. Usability, trustworthiness, perceived clinical value, and visual interference will each be rated on a five-point Likert-type scale and reported separately. Each domain has a minimum score of 1 and a maximum score of 5. For usability, trustworthiness, and perceived clinical value, 1 indicates very low and 5 indicates very high. Higher scores indicate more favorable ratings. For visual interference, 1 indicates no interference and 5 indicates very severe interference. Higher scores indicate greater visual interference.
NASA-TLX score
Surgeons' perceived workload will be assessed after each evaluated procedure using the original weighted National Aeronautics and Space Administration Task Load Index (NASA-TLX). Six dimensions are rated from 0 to 100: mental demand, physical demand, temporal demand, performance, effort, and frustration. Fifteen pairwise comparisons determine each dimension's weight, ranging from 0 to 5; the weights sum to 15. The overall score is calculated as the sum of each dimension's rating multiplied by its weight, divided by 15. The total score ranges from 0 to 100; higher scores indicate greater perceived workload and lower scores indicate less perceived workload.
Performance in human-AI comparison
The artificial intelligence system and human annotators will identify anatomical structures on the same set of 100 to 300 challenging intraoperative images randomly sampled from the external validation set. Their outputs will be compared with expert-reviewed reference annotations. Accuracy is the percentage of correct determinations. Sensitivity is the percentage of reference-positive determinations correctly identified, and specificity is the percentage of reference-negative determinations correctly identified. Each metric will be reported separately for the artificial intelligence system and for three human groups: experts (more than 100 relevant procedures), intermediate-experience physicians (50 to 100 relevant procedures), and novices (residents, students, or surgical assistants).
Time to correct anatomical structure identification in surgical videos: human-AI comparison
Time to correct anatomical identification will be compared among the AI system, expert surgeons, intermediate-experience surgeons, and novice participants using the same surgical video clips from the external test set. Time will be measured from each target structure's first identifiable appearance, predefined by independent experts, until its location is correctly indicated. Human participants will indicate the structure by clicking or pointing; AI identification will be recorded when the system first correctly indicates the structure's location. Clips will be presented at normal speed without pausing or replay, and AI will process frames online, including processing latency. Results will be expressed in seconds per target structure for each group. Targets not identified by clip end will be right-censored in time-to-event analyses. Lower values indicate faster identification.
Locations
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Chinese PLA General Hospital
Lead sponsor
Tongji Hospital
Collaborator
Tianjin Medical University Second Hospital
Collaborator
Peking University First Hospital
Collaborator
West China Hospital
Collaborator
First Affiliated Hospital, Sun Yat-Sen University
Collaborator
Sun Yat-Sen Memorial Hospital of Sun Yat-Sen University
Collaborator
RenJi Hospital
Collaborator
The Second Affiliated Hospital of Xi 'an Jiaotong University
Collaborator
This trial is not recruiting at the moment. You can still explore other options: