About this trial
Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Presenting for a routine eye exam
≥ 18 years of age
Able and willing to give consent
Disqualifiers
History of corneal scarring
Active infection of the eye
Ocular surface trauma or infection
Trial design
Parallel
Treatments tested in this trial
Standard GAT
DeviceStandard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice
Fixed force GAT
DeviceAn investigational device similar to standard GAT
Upright applanating prototype
DeviceAn investigational device, prism used in standard and fixed-force GAT is attached to a portable device
Pneumotonometer
DeviceThis is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure
Treatment groups
Trial outcomes
Primary outcomes
Intraocular pressure (IOP) as measured by standard GAT
this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.
Intraocular pressure (IOP) as measured by fixed force GAT
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular.
Intraocular pressure as measured by upright applanating
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.
Intraocular pressure as measured by supine Pneumotonometer