About this trial
The purpose of this clinical investigation is to evaluate the clinical performance, safety, and clinical benefit of the BrainProtect System in reducing early postoperative neurological complications in adult participants undergoing elective coronary artery bypass grafting (CABG) with cardiopulmonary bypass (CPB).
The investigation aims to verify the claim that real-time detection and correction of cerebral autoregulation (CA) impairment during CPB improves neurological outcomes compared to standard perioperative management. The study is designed to demonstrate superiority of BrainProtect-guided management over standard care.
This is a Multicenter, two-arm, blinded, randomized controlled pivotal (confirmatory) clinical investigation of a non-CE marked investigational medical device. Adult participants undergoing elective coronary artery bypass graft (CABG) surgery requiring cardiopulmonary bypass (CPB) will be randomized in a 1:1 ratio to either the investigational group or the control group.
In the investigational group, standard perioperative management during CABG with CPB will be performed. Real-time cerebral autoregulation (CA) monitoring using the BrainProtect System will be applied during CPB.
In the control group, standard perioperative management during CABG with CPB will be performed. Conventional systemic hemodynamic monitoring will be used. The BrainProtect System will not be used, and no real-time cerebral autoregulation monitoring data will be available to the clinical team.
Postoperative neurological outcomes (POCD and POD) will be assessed in both groups using predefined standardized neuropsychological tests and CAM-ICU assessments during the early postoperative period. In addition, the length of the index hospitalisation and occurrence of hospital readmission within 31 days after surgery and the reason for readmission will be recorded.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Age 18-69 years.
Patient has provided written informed consent to participate in the study.
Qualification for the elective coronary artery bypass graft (CABG) surgery requiring CPB.
Disqualifiers
Female subjects who are pregnant or breastfeeding.
Patients with an eye disease that may affect monitoring results, as determined by the Investigator.
Patient who had an orbital or eye injury.
Patients who have acute or chronic but currently exacerbated respiratory disease.
Trial design
Parallel
Treatments tested in this trial
BrainProtect System
DeviceThe BrainProtect System consists of the Archimedes 02 monitoring system, the BrainProtect Plug-in Modulator, and BrainProtect Software. The Archimedes 02 monitoring system is positioned on the participant's closed eyelids to enable non-invasive monitoring of intracranial pressure pulse waves and cerebral autoregulation-related signals. Following initiation of cardiopulmonary bypass and confirmation of stable extracorporeal circulation, the BrainProtect Plug-in Modulator is activated. Rectangular flow modulation is initiated and maintained during cardiopulmonary bypass unless clinically contraindicated. The perfusionist has real-time access to cerebral autoregulation monitoring data. When predefined algorithm-based criteria for impaired cerebral autoregulation are detected, a visual event is generated. Blood pressure management decisions are made by the perfusionist based on cerebral autoregulation status and standard clinical practice.
Standard perioperative management
Procedure/SurgeryStandard perioperative management during CABG with CPB will be performed. Conventional systemic hemodynamic monitoring will be used. Adjunctive cerebral oxygenation monitoring using near-infrared spectroscopy (NIRS) may be used if routinely applied at the site.
Treatment groups
Trial outcomes
Primary outcomes
Number of Participants With Postoperative Delirium and/or Postoperative Cognitive Dysfunction (Composite Neurological Outcome)
Number (percentage) of participants with a positive binary composite outcome of postoperative delirium (POD) and/or postoperative cognitive dysfunction (POCD). Individual assessments are used only to classify participants as composite-positive or composite-negative. Individual test scores are not reported in this outcome measure. POD: ≥1 positive Confusion Assessment Method for the ICU (CAM-ICU) assessment on postoperative Days 1-4 (Feature 1 acute onset/fluctuating course and Feature 2 inattention, plus Feature 3 altered consciousness or Feature 4 disorganized thinking). POCD: decline from preoperative baseline (1-2 days before surgery) to postoperative Days 4-7 in ≥2 of 4 tests: Mini-Mental State Examination (≥3-point decrease), Hopkins Verbal Learning Test-Revised (≥20% decrease), Digit Symbol Substitution Test (≥20% decrease), Schulte Table test (≥20% increase in completion time). A participant is composite-positive if POD and/or POCD is present.
Secondary outcomes
Incidence of Postoperative Delirium (POD)
Incidence of postoperative delirium (POD), defined as at least one positive CAM-ICU assessment during the postoperative assessment period.
Incidence of Postoperative Cognitive Dysfunction (POCD)
Incidence of postoperative cognitive dysfunction (POCD), defined as a decline from baseline performance in at least two tests within the neuropsychological test battery consisting of MMSE, HVLT-R, DSST, and Schulte Table test.
Odds ratio for the composite neurological outcome according to duration of the longest cerebral autoregulation impairment event
The odds ratio for the composite neurological outcome according to the duration of the longest intraoperative cerebral autoregulation (CA) impairment event per participant will be estimated, to evaluate whether longer CA impairment events are associated with a higher risk of the composite outcome. CA impairment event duration will be measured in seconds using the BrainProtect System during CPB. The composite neurological outcome is binary (POD and/or POCD present vs absent). The odds ratio with confidence interval will be estimated using exploratory logistic regression. This analysis is supported by previous findings showing an association between longer single CA impairment events and postoperative neurological outcomes (Kasputyte et al., 2025)
Odds ratio for Postoperative Delirium according to duration of the longest cerebral autoregulation impairment event
The odds ratio for postoperative delirium (POD) according to the duration of the longest intraoperative cerebral autoregulation (CA) impairment event per participant will be estimated, to evaluate whether longer CA impairment events are associated with a higher risk of POD. CA impairment event duration will be measured in seconds using the BrainProtect System during CPB. POD is a binary outcome (present/absent) assessed daily on postoperative Days 1-4 using CAM-ICU; POD is present if ≥1 CAM-ICU assessment is positive. The odds ratio with confidence interval will be estimated using exploratory logistic regression. This analysis is supported by previous findings linking longer CA impairment duration with POD (Kasputyte et al., 2025).
Other outcomes
Correlation Coefficient Between Intracranial Pressure Pulse Waveforms Recorded During Intact and Impaired Cerebral Autoregulation
Intracranial pressure (ICP) pulse waves will be recorded non-invasively by the BrainProtect System during cardiopulmonary bypass and classified according to cerebral autoregulation (CA) status as intact or impaired. For each participant, representative ICP pulse waveforms during intact and impaired CA periods will be compared as complete waveform shapes. Waveform morphology similarity will be quantified using a correlation coefficient (r), reported as a dimensionless value. Lower correlation will indicate greater change in ICP pulse-wave morphology between intact and impaired CA states.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Scienceforbrain, UAB
Lead sponsor
KCRI
Collaborator
This trial is not recruiting at the moment. You can still explore other options: