Clinical Trial for the Validation of AR Based Neuronavigation System

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorSeoul National University Hospital

About this trial

The goal of this clinical trial is to test augmented reality (AR) based neuronavigation system in surgeries for patients of brain neoplasm or cerebral vascular disease. The main questions it aims to answer are:

• AR based neuronavigation system can achieve accuracy that is not inferior to conventional intraoperative navigation system.

Participants will participate the study after informed consent. When participants undergo surgery for their brain tumor, we will set up 2 types of neuronavigation, conventional navigation system and developed AR based neuronavigation system. Surgeon will plan and conduct surgery based on only conventional navigation system, but 3D errors at several selected points between two types of navigation will be measured and analyzed.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Adults aged 18 years or older with a brain tumor or cerebrovascular disease who have undergone preoperative CT

Scheduled for surgical treatment of the brain tumor or cerebrovascular disease

Surgical position allowing exposure and measurement of bilateral Kocher's points and the glabella without significant head rotation

Disqualifiers

Patient or legal representative does not consent to participation, including consent to intraoperative video recording

Anatomical deformity of the skull or facial region due to previous surgery or other causes that precludes identification of Kocher's points or the glabella

Emergency surgery

Surgical position (e.g., prone, lateral rotation of 90° or more) that does not allow simultaneous exposure of bilateral Kocher's points and the glabella

Trial design

Design model

Single group

Treatments tested in this trial

  • AR-based navigation

    Device

    Patient-specific 3D models are generated from preoperative CT using MEDIP PRO and displayed on a tablet (iPad) via MEDIP PRO AR, superimposed on the patient using markerless surface registration. Registration and TRE measurement are video-recorded for retrospective analysis of registration time. The device is used for accuracy evaluation only and does not guide any surgical step.

Treatment groups

80 Participants
are divided into 1 treatment group
Group A: Single armExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Target Registration Error (TRE) of AR-based navigation at anatomical landmarks

Linear distance (mm) between the landmark position displayed by MEDIP PRO AR and the actual position on the patient's skin/skull surface, measured directly with a ruler at three fixed landmarks per participant (right Kocher's point, left Kocher's point, glabella). Measured immediately after registration (pre-drape). Reported as mean ± SD or median (IQR) overall and per landmark.

Time frame
Intraoperative, immediately after AR registration before surgical draping on the day of surgery
2

Registration time of AR-based navigation

Time (minutes, seconds) from AR software launch to completion of registration, determined by retrospective analysis of intraoperative video recording. Reported separately for manual and automatic registration.

Time frame
Intraoperative, on the day of surgery

Secondary outcomes

1

Post-drape TRE after QR marker-based registration restoration

TRE (mm) at the same three landmarks measured after surgical draping and restoration of registration using QR markers attached to the skin.

Time frame
Intraoperative, immediately after draping, on the day of surgery
2

Incidence of registration drift after draping

Proportion of landmarks showing visible displacement of the AR overlay of ≥7.5 mm after draping, assessed on a second tablet held in a fixed mount.

Time frame
Intraoperative, immediately after draping, on the day of surgery
3

Adverse events and device malfunctions

Number and description of adverse events and device errors occurring during AR navigation application.

Time frame
Intraoperative through end of surgery, on the day of surgery
4

One-step processing time from CT acquisition to 3D model delivery (SNUH only)

Total elapsed time from CT image receipt on the server to delivery of the 3D mesh to the tablet, including automatic segmentation, calculated from system log timestamps. Success of automatic segmentation (no manual correction required) is also recorded.

Time frame
Preoperative, within 24 hours before surgery

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.