Clinical Validation of the Metrix CTNGTV Test

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexFemale
Age14+
SponsorAptitude Medical Systems

About this trial

The Metrix CT/NG/TV Test will be evaluated for use in Non-Laboratory settings in a home testing environment utilizing the clinical study design described herein. The study will take place in simulated home environments which will be set up within or near active clinical settings (e.g., urgent care facilities). This will be a prospective study conducted at three or more investigational sites located within the United States for the clinical validation of the Metrix CT/NG/TV Test for the detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis from vaginal swab samples. Additional sites may be added to the study in order to meet minimum subject/sample enrollment requirements and geographic prevalence of STIs. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test. The patient infectious status will be determined with two separate FDA-cleared assays for the detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.

Eligibility criteria

This trial accepts healthy volunteers

Qualifiers

1. Subject is willing and able to give voluntary written informed consent (and assent, if under age 18) before any study-related procedure is performed.

2. Subject was assigned female sex at birth. (Pregnant and breastfeeding subjects are eligible.)

3. Subject is aged 14 years or older at the time of enrollment.

4. Subject is sexually active within the last 6 months.

Disqualifiers

1. Subject does not understand or is not able and willing to sign the study informed consent and/or assent.

2. Subject or guardian is not able to comply with swab collection requirements.

3. Subject has a medical condition, serious illness, or other circumstance that, in the Investigator's judgment, could jeopardize the subject's safety or interfere with study procedures.

4. Subject has used any of the following in the genital area within 48 hours prior to enrollment: antiperspirants, deodorants, or the following vaginal products: douches, washes, lubricants, vaginal wipes, vaginal moisturizers, or feminine hygiene spray.

Trial design

Design model

Single group

Treatments tested in this trial

  • Metrix CT/NG/TV Test

    Device

    The Metrix CT/NG/TV Test is a real-time reverse transcription loop-mediated isothermal amplification reaction (RT-LAMP) test that has been designed to detect DNA from Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis via electrochemical detection using the Metrix Reader. Detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis gene targets will produce a positive result, differentiated by each target. If no target is detected and the internal control amplifies, the result is negative. If no amplification is detected in any of the channels, an invalid result is displayed.

Treatment groups

2,000 Participants
are divided into 1 treatment group
Group A: OTC StudyExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Qualitative Detection of DNA from Chlamydia trachomatis, Neisseria gonorrhoeae and Trichomonas vaginalis, and Measurement of Positive Percent Agreement(PPA)/Sensitivity and Negative Percent Agreement (NPA)/Specificity with FDA-cleared tests

Qualitative Detection of DNA from Chlamydia trachomatis, Neisseria gonorrhoeae and Trichomonas vaginalis in unprocessed, self-collected vaginal swab will be collected throughout the study. Primary objective of this study is to evaluate the performance (positive percent agreement/sensitivity, negative percent agreement/specificity, and 95% confidence intervals for upper and lower bounds) of the Metrix CT/NG/TV Test. Comparator testing will be performed using at least 2 FDA-cleared assays to determine the status of each sample for comparison to results produced by the candidate test. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test.

Time frame
30 minutes

Secondary outcomes

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.