The purpose of this research, which has been determined as non-significant risk by the central IRB overseeing the study, is to obtain information to help further develop a machine (a medical device) to measure the pressure around the brain from the outside (this pressure is called intracranial pressure or ICP). Monitoring and managing ICP is an important part of care for patients with conditions such as Traumatic Brain Injury (TBI). However, the current way of measuring ICP requires surgery to drill a hole into the skull, and therefore can introduce additional risks such as infections and pain.
Recent research has shown it may be possible to measure ICP without needing surgery. This technology is in development, but large amounts of data is required to build these new devices.
Through collecting a large database of information from patients who have both the routine surgical device and the research device applied to their head, the research team will work to develop and test an effective and potentially safer way of monitoring patient ICP.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Male or female sex at birth, and aged 18 years or older on the date of enrollment.
Receiving continuous invasive ICP monitoring (Bolt or EVD) as part of standard care.
Invasive ICP monitor catheter penetrating the parenchyma or ventricles.
Receiving continuous invasive ABP monitoring as part of standard care.
Disqualifiers
Presence of any implant (cosmetic or otherwise) in the frontal bone in such proximity to the CoMind One EFS Sensor that they might physically touch.
Open wounds on the forehead such that CoMind One EFS Sensor cannot be safely placed over an area of intact skin
Presenting with radiographic evidence of a non-intact skull at the recording site on admission.
If patient is enrolled in an intervention/study that may interfere with SoC ICP measurements or the CoMind One EFS device measurement then the patient is ineligible.
Trial population
Any participant (18 years or older) who is undergoing invasive ICP monitoring and continuous ABP monitoring as part of their normal medical treatment at selected clinical sites. The majority of such participants will have sustained a Traumatic Brain Injury (TBI), but some may have Intracerebral Haemorrhage (ICH), Encephalitis, various metabolic Encephalopathies (notably hepatic), Hydrocephalus (both communicating and non-communicating), or Stroke, although this is not an exhaustive list of participant conditions.
Trial design
Case-only
Prospective
Treatments tested in this trial
Not listed
Trial groups
Trial outcomes
Primary outcomes
A test of the difference in estimation errors between two non-invasive ICP estimation models: one trained with ABP and CoMind One EFS CBFi, and one trained only using ABP
Models trained using CoMind CBFi and ABP will be evaluated based on their limits of agreement (LOA) with invasive ICP.
The threshold limit of agreement will be evaluated at three levels corresponding to a model with clinical utility that is optimal, high, or moderate clinical utility. These levels of clinical utility are defined as follows: ● Optimal clinical utility: a model trained using CoMind CBFi and ABP can achieve LOA with invasive ICP less-than or equal to +/-2 mmHg in the range of ICP between 10-30 mmHg, and less-than or equal to +/-4 mmHg in the ranges of 0-10 mmHg, and 30-50 mmHg. ● High clinical utility: a model trained using CoMind CBFi and ABP can achieve LOA with invasive ICP less-than or equal to +/-3 mmHg in the range of ICP between 10-30 mmHg, and less-than or equal to +/-5 mmHg in the ranges of 0-10 mmHg, and 30-50 mmHg. ● Moderate clinical utility: a model that achieves LOA with invasive ICP less-than or equal to +/-4 mmHg in the range of ICP between 10-30 mmHg, and less-than or equal to +/-6 mmHg in the ranges of 0-10 mmHg, and 30-50 mmHg.
Secondary outcomes
Demonstration that non-invasive ICP estimation using the CoMind One EFS device is insensitive to differences in skin tone across the population.
Demonstration of a non-invasive ICP model that can classify the transition between ICP 'states', i.e. transitions between low ICP (less than 20 mmHg), and high ICP (greater than 20 mmHg).
Demonstration of a non-invasive ICP waveform prediction model and assessment of its accuracy against the invasively-measured ICP waveform.
Demonstration of an non-invasive ICP model that can predict trends in ICP
Other outcomes
A comparison of metrics of CAR derived from CoMind One EFS CBFi, CoMind One EFS device estimates of ICP, and ABP.