Correlation Between Visual Reading and DOlab Software Quantitative Analysis of DaT-SPECT Images

Trial statusRecruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
Age18+
SponsorDorian Technologies S.r.l.

About this trial

The goal of this study is to find out whether the quantitative results of DOlab software (DaT-SPECT workflow) agree with the visual reading of brain scans performed by nuclear medicine physicians. The scans are 123I-ioflupane DaT-SPECT images of adults who were evaluated for dopamine transporter dysfunction, a condition seen in Parkinson's disease, other parkinsonian syndromes and dementia with Lewy bodies.

The main questions it aims to answer are:

* Do the visual scores given by expert readers (0 = normal to 3 = marked abnormality) correlate with the values calculated by the software? * Does using the software report change or refine the readers' classification compared with visual reading alone? * How usable do readers find the software?

The study uses existing scans of 120 patients collected between January 2015 and May 2025 at two Italian hospitals. No new tests or treatments are given to participants, and the software results are not used for their clinical care.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Age 18 years or older.

Referred for evaluation of nigrostriatal dopaminergic dysfunction to one of the participating centers between January 2015 and May 2025.

Informed consent for data collection, or documentation that all reasonable efforts to obtain consent were made (Italian Privacy Code, art. 110).

were acquired according to the PPMI imaging protocol;

Disqualifiers

Age under 18 years.

Use of drugs or substances that may significantly affect DaT binding (e.g., cocaine, amphetamines, CNS stimulants, opioid derivatives, antidepressants, adrenergic agonists, anticholinergics, anesthetics) without a washout of at least five biological half-lives before DaT-SPECT, per the EANM/SNMMI guidelines.

DaT-SPECT performed outside the January 2015 - May 2025 period.

Images not meeting the acquisition and quality specifications listed above.

Trial population

Adults referred for evaluation of nigrostriatal dopaminergic dysfunction who underwent 123I-ioflupane DaT-SPECT between January 2015 and May 2025 at Santo Stefano Hospital (Prato) or the University Hospital of Padua, Italy. At each center, 15 consecutive scans per visual grade (negative, mild, moderate and marked abnormality) are selected retrospectively.

Trial design

Design model

Case-only

Time perspective

Retrospective

Treatments tested in this trial

  • DOlab software - DaT-SPECT workflow

    Device

    Investigational software (not CE-marked, class IIa) that automatically quantifies striatal 123I-ioflupane binding using four metrics (SBR, Southampton SBR, Basal Ganglia V3, Volume Fraction) and compares them with a PPMI-based normative database. It is applied retrospectively to existing images, and its output is not used for patient care.

  • Visual interpretation (comparator)

    Diagnostic test

    Visual reading of the same 123I-ioflupane images by blinded nuclear medicine physicians, scored 0-3 for the caudate, anterior putamen and posterior putamen of each hemisphere, according to the EANM/SNMMI guidelines.

Treatment groups

120 Participants
are divided into 1 treatment group
Group A: Adults referred for DaT-SPECT2 interventions

Trial outcomes

Primary outcomes

1

Correlation between visual score and DOlab quantitative metrics

Pearson correlation coefficient (with 95% CI) between the readers' visual score (ordinal 0-3; 0 = normal, 3 = marked abnormality) and each continuous metric produced by DOlab software (SBR, Southampton SBR, Basal Ganglia V3, Volume Fraction) for the caudate, anterior putamen and posterior putamen of each hemisphere. If the Pearson assumptions are not met, the Spearman rank correlation is reported.

Time frame
Phase 1 reading session, within 4 months of study start

Secondary outcomes

1

Agreement between visual and software-supported semi-quantitative classification

Cohen's kappa coefficient between the classifications given by the same readers with visual reading alone (Phase 1) and with the DOlab software report (Phase 2).

Time frame
Phase 2 reading session, within 4 months of study start
2

Number of scans reclassified with software support

Number and percentage of scans whose classification changes between visual reading alone and software-supported reading (e.g., borderline to abnormal), tested with McNemar's test for paired proportions.

Time frame
Phase 2 reading session, within 4 months of study start
3

Software usability (System Usability Scale)

SUS questionnaire completed by the 4 readers: 10 items on a 5-point scale. SUS total score from 0 to 100, where higher scores indicate better perceived usability; item responses are also reported as median and IQR.

Time frame
After the Phase 2 reading session, within 4 months of study start

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Dorian Technologies S.r.l.

Lead sponsor

JSB Solutions S.R.L.

Collaborator