Determination of Specificity and Sensitivity in Detecting Myocardial Infarction Using the EDE CARDIO BIO-MAP Machine

Trial statusNot yet recruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
Age18+
SponsorWeill Cornell Medical College in Qatar

About this trial

The goal of this observational study is to learn whether a non-invasive electromagnetic wave scanner (EDE CARDIO BIO-MAP) can help diagnose myocardial infarction (heart attack) in adults (18 years and older) who come to the emergency department with chest pain and suspected myocardial infarction. The main questions it aims to answer are:

* How sensitive and specific is the EDE scanner in diagnosing myocardial infarction (MI) compared with standard blood tests (e.g., cardiac troponin)? * How sensitive and specific are the immediate blood measurements from a portable lab device compared with conventional hospital laboratory measurements? Researchers will compare the results of the EDE scanner and the portable lab with standard hospital blood test results and the final clinical diagnosis to see how accurately each detects MI.

Participants will:

* Have a research visit (30 minutes to 1 hour) during their emergency care, including a quick scan with the Exponential Deep Examination (EDE) device (attached to a phone or tablet) and a small research blood sample tested on a portable lab device * Repeat the scan and blood sample 4 to 6 hours later, only if a second blood draw is already needed for their clinical care * Answer questions about their medical and social history, and allow the study team to collect relevant results from their medical record Scanner results will not be shared with participants or used for their clinical care.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Age ≥18 years.

Adults presenting to the emergency room for acute chest pain.

English or Arabic speakers.

Disqualifiers

Pregnancy (self-reported).

Adults with cognitive disorders.

Unconscious patients.

Citizens from countries where the General Data Protection Regulation (GDPR) applies, due to regulatory constraints.

Trial population

Participants will be recruited at Hamad Medical Corporation

Trial design

Design model

Cohort

Time perspective

Prospective

Treatments tested in this trial

Not listed

Trial groups

No trial groups listed

Trial outcomes

Primary outcomes

1

Sensitivity and specificity of the EDE CARDIO BIO-MAP scanner(in vivo) for detecting myocardial infarction

Sensitivity (true positives / \[true positives + false negatives\]) and specificity (true negatives / \[true negatives + false positives\]) of the scanner in emergency room patients with suspected myocardial infarction, compared with standard hospital laboratory results (cardiac troponin, creatine kinase-myocardial band, myoglobin, B-type natriuretic peptide, N-terminal pro-B-type natriuretic peptide) and the final clinical diagnosis in the medical record.

Time frame
Baseline

Secondary outcomes

1

Sensitivity and specificity of the portable lab's immediate blood measurements (in vitro)

Sensitivity and specificity of the portable lab's immediate blood measurements for myocardial infarction, compared with conventional hospital laboratory measurements, using a confusion matrix.

Time frame
Baseline
2

Accuracy of the EDE scanner

Accuracy = (true positives + true negatives) / all results;

Time frame
Baseline
3

Misclassification rate of the EDE scanner

Misclassification rate = (false positives + false negatives) / all results;

Time frame
Baseline
4

Precision of the EDE scanner

precision = true positives / (true positives + false positives)

Time frame
Baseline

Other outcomes

1

Change in cardiac biomarker levels between initial assessment and 4-6 hour follow-up

Change in cardiac biomarker levels( Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB) between the initial assessment and the follow-up, in participants who have a second clinical blood draw.

Time frame
Baseline and 4-6 hours after initial assessment
2

Diagnostic performance of the scanner by sex

Subgroup analysis of scanner sensitivity and specificity by sex

Time frame
Baseline
3

Diagnostic performance of the scanner by age group

Subgroup analysis of scanner sensitivity and specificity by age group

Time frame
Baseline
4

Diagnostic performance of the scanner by symptom profile

Subgroup analysis of scanner sensitivity and specificity by symptom profile.

Time frame
Baseline

Locations

This trial has no locations.

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Weill Cornell Medical College in Qatar

Lead sponsor

Hamad Medical Corporation

Collaborator

This trial is not recruiting at the moment. You can still explore other options: