About this trial
The investigators are developing a video laryngoscope equipped with artificial intelligence technology to enable automatic identification of key airway anatomical structures and real-time guidance during intubation, with the aim of improving the success rate and efficiency of endotracheal intubation in emergency settings.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Patient inclusion criteria: Adult patients in critical condition requiring endotracheal intubation via laryngoscopy (age ≥ 18 years).
Disqualifiers
pregnant women, prisoners, recent history of airway surgery, oral or pharyngeal malformations, inability to perform randomization due to the need for emergency endotracheal intubation, or when the investigator deemed it necessary or contraindicated to use AIVL, VL, or DL during the first intubation attempt.
Trial design
Parallel
Treatments tested in this trial
AI-video laryngoscopy (AIVL)
DeviceWhen using the AIVL for the first attempt at laryngoscopy intubation, the operator shall receive pre-procedural training, which primarily covers the functional buttons, the prompt interface, and image display.
Conventional Video Laryngoscope (VL)
DeviceA conventional video laryngoscope will be used for the first endotracheal intubation attempt. The device displays the airway image without AI-assisted identification of anatomical structures or real-time AI intubation guidance.
Treatment groups
Trial outcomes
Primary outcomes
First-pass intubation success rate
Proportion of participants with successful endotracheal tube placement on the first laryngoscopic attempt, confirmed by end-tidal carbon dioxide monitoring.
Secondary outcomes
Intubation time
Time in seconds from insertion of the laryngoscope into the oral cavity until the endotracheal tube passes through the glottis.
Severe complications occurring within 2 minutes after intubation were defined as severe hypoxemia (peripheral SpO₂ <80%), severe hypotension (systolic blood pressure <65 mmHg), initiation of vasoactive agents, cardiac arrest, or death.
AI recall
True-positive anatomical detections divided by true-positive plus false-negative detections, assessed against reference annotations.
Duration of mechanical ventilation
Duration of mechanical ventilation, measured in days.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Second Affiliated Hospital, Zhejiang University, School of Medicine
Lead sponsor
Shaoxing Central Hospital
Collaborator