About this trial
Study 1 (S1). In a range of adult patients with sNF and controls, there is no significant difference between the POC-ANC and laboratory analyser.
The diagnostic study will be performed 24 hours a day. Informed consent will be sought from every patient for a finger prick (capillary blood) and venous blood sample.
Briefly, paired capillary (finger prick) and venous EDTA blood samples will be obtained from 317 patients meeting the inclusion criteria.
Prospective diagnostic accuracy study. The target accuracy index is the lower 95%CI of sensitivity, although we evaluate other important factors of diagnostic accuracy (see data analysis below).
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
≥18 years; AND
Suspected low ANC; AND
Stable.
Disqualifiers
No consent obtained
Blood sampling my hinder the management of the patient according to the doctor or nurse.
Trial design
Single group
Treatments tested in this trial
Hemocue® WBC DIFF system
Diagnostic testBriefly, paired capillary (finger prick) and venous EDTA blood samples will be obtained from 317 patients meeting the inclusion criteria. Reference counts (Gold standard) will be obtained by a standardised Sysmex XE-5000 automated cell counter (Sysmex Corporation, Kobe, Japan). All samples will be analysed in duplicates on the HemoCue WBC DIFF system and in single replicate on the Sysmex XE-5000 within 8 hours after collection. The timespan between analyses on both systems should be less than 2 hours for each sample.
Treatment groups
Trial outcomes
Primary outcomes
sensitivity of POC-ANC compared with the laboratory reference method
The target accuracy index is the lower 95%CI of sensitivity, although we evaluate other important factors of diagnostic accuracy (see data analysis below). The primary analysis is on sensitivity of 97 - 99.9% looking to safely rule out NF.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
This trial is not recruiting at the moment. You can still explore other options: