Diagnostic Accuracy, Clinical Utility and Implementation of Point-of-Care Absolute Neutrophil Count in Suspected Neutropenic Fever (DUCATIS) - Diagnostic Accuracy (S1)

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorThe University of Hong Kong

About this trial

Study 1 (S1). In a range of adult patients with sNF and controls, there is no significant difference between the POC-ANC and laboratory analyser.

The diagnostic study will be performed 24 hours a day. Informed consent will be sought from every patient for a finger prick (capillary blood) and venous blood sample.

Briefly, paired capillary (finger prick) and venous EDTA blood samples will be obtained from 317 patients meeting the inclusion criteria.

Prospective diagnostic accuracy study. The target accuracy index is the lower 95%CI of sensitivity, although we evaluate other important factors of diagnostic accuracy (see data analysis below).

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

≥18 years; AND

Suspected low ANC; AND

Stable.

Disqualifiers

No consent obtained

Blood sampling my hinder the management of the patient according to the doctor or nurse.

Trial design

Design model

Single group

Treatments tested in this trial

  • Hemocue® WBC DIFF system

    Diagnostic test

    Briefly, paired capillary (finger prick) and venous EDTA blood samples will be obtained from 317 patients meeting the inclusion criteria. Reference counts (Gold standard) will be obtained by a standardised Sysmex XE-5000 automated cell counter (Sysmex Corporation, Kobe, Japan). All samples will be analysed in duplicates on the HemoCue WBC DIFF system and in single replicate on the Sysmex XE-5000 within 8 hours after collection. The timespan between analyses on both systems should be less than 2 hours for each sample.

Treatment groups

317 Participants
are divided into 1 treatment group
Group A: Consented participantsExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

sensitivity of POC-ANC compared with the laboratory reference method

The target accuracy index is the lower 95%CI of sensitivity, although we evaluate other important factors of diagnostic accuracy (see data analysis below). The primary analysis is on sensitivity of 97 - 99.9% looking to safely rule out NF.

Time frame
All samples will be analysed in duplicates on the HemoCue WBC DIFF system and in single replicate on the Sysmex XE-5000 within 8 hours after collection.

Secondary outcomes

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

This trial is not recruiting at the moment. You can still explore other options: