The purpose of the study is to assess precision of the GE Lunar iDXA, Hologic Horizon W DXA scanner, and the Stratec peripheral quantitative computed tomography (pQCT) scanner.
Aim 1: To assess standard deviation (SD) and coefficient of variation (CV) of testing bone mineral, bone geometry, and soft tissues in inanimate objects and adults. Tests on adults will be structured to assess:
1. Within- and between-day variability of testing 2. Within- and between-technician variability of testing 3. Variability due to positioning and post-scan processing 4. Variability due to different DXA scanning equipment
Aim 2: To assess the least significant change (LSC) for testing bone mineral and soft tissues in adults.
Aim 3: To assess strengths and limitations of DXA and pQCT testing, including sensitivity to movement, rate of re-rescanning, etc. which are crucial components for demonstrating feasibility in grant applications.
Aim 4: To construct a database of DXA and pQCT test results and quality control procedures, showing our laboratory's quality control level.
Eligibility criteria
This trial accepts healthy volunteersQualifiers
None
Disqualifiers
for females, being pregnant
Having had X-ray procedures using contrast material in the previous 3 days
Having medical devices that interfere with scan accuracy
Wearing external metal objects that cannot be removed
Trial population
Volunteers will include men and women of all ages. Women must not be pregnant.
Trial design
Cohort
Cross-sectional
Treatments tested in this trial
DXA and pQCT precision testing
DeviceThe participants are only being observed; there is no intervention other than DXA and pQCT testing for the purposes of looking at instrument/technician precision..
Treatment groups
Trial outcomes
Primary outcomes
Precision of iDXA BMD
Standard Deviation and coefficient of variation for BMD measurements obtained on the GE Lunar iDXA
Precision of Hologic Horizon W DXA BMD
Standard Deviation and coefficient of variation for BMD measurements obtained on the Hologic Horizon W
Precision of iDXA Body Composition
Standard Deviation and coefficient of variation for body composition measurements obtained on the GE Lunar iDXA
Precision of Hologic Horizon W DXA Body Composition
Standard Deviation and coefficient of variation for body composition measurements obtained on the Hologic Horizon W
Secondary outcomes
LSC for the iDXA BMD
Least significant change for BMD using the GE Lunar iDXA
LSC for the Hologic Horizon W BMD
Least significant change for BMD using the Hologic Horizon W DXA
LSC for the iDXA body composition
Least significant change for body composition using the GE Lunar iDXA
LSC for the Hologic Horizon W body composition
Least significant change for body composition using the Hologic Horizon W DXA