About this trial
The proposed study is an investigator-initiated study that aims to measure the accuracy of a wearable seizure detection and prediction device (Ear-Seizure Detection Device (EarSD)) by simultaneous recording with conventional video-EEG (Electroencephalogram) on patients with epileptic seizures in the Epilepsy Monitoring Unit of the hospital.
Eligibility criteria
This trial accepts healthy volunteersQualifiers
Age ≥ 18 years.
Patients admitted to UMass Memorial Epilepsy Monitoring Unit (EMU) for long term video-EEG monitoring as part of standard care of both focal and generalized epilepsy.
Willing to wear the wearable device.
Ability to provide informed consent
Disqualifiers
Subjects wearing other ear devices such as hearing aids.
Inability or unwillingness to provide informed consent.
Irritation of the skin where the device is to be placed.
Patients with intracranial electrodes placement.
Trial design
Single group
Treatments tested in this trial
Ear-SD
DeviceThe Ear-SD is a purely EEG recording device Continuous Electroencephalogram (cEEG), Electromyogram (EMG), Electrooculogram (EOG), Photoplethysmogram (PPG), Electrodermoactivity (EDA), and Inertial Measurement Unit (IMU). The Ear-SD device rests on the ears and connects to the scalp by two sticker electrodes.
Electroencephalogram
Diagnostic testStandard 21-channel scalp-continuous electroencephalogram (cEEG) with video recording and electrocardiogram (ECG)
Treatment groups
Trial outcomes
Primary outcomes
Seizure Recording Criteria 1
Recordings of Bioelectrical signal of subjects with the wearable device and simultaneous continuous EEG data is collected for the duration of hospitalization of participants. Outcome measures reported include number of seizure events per participant.
Seizure Recording Criteria 2
Recordings of Bioelectrical signal of subjects with the wearable device and simultaneous continuous EEG data is collected for the duration of hospitalization of participants. Outcome measures reported include average duration of each seizure in minutes and seconds and total recording time in hours aggregated to arrive at one reported value seizure classification.
Seizure Recording Criteria 3
Recordings of Bioelectrical signal of subjects with the wearable device and simultaneous continuous EEG data is collected for the duration of hospitalization of participants. Outcome measures reported include reported value seizure classification. Seizure classification includes Unclassified (UC), Focal Onset Aware (FOA), Focal Onset Impaired (FOIA), Focal to Bilateral Tonic-Clonic (FBTC).
Data Interpretation
EarSD extracted EEG signals from the log file plotted alongside EDF files from cEEG are measured and compared to detect seizure onset and offset times for data interpretation. Two-minute segments of cEEG European Data Format (EDF) consisting of non-seizure signals from periods before and after the seizures, and non-seizure signals from periods of daily activities like talking, eating, and walking are involved in the comparison to detect seizure onset and offset times. Prediction measurement of Seizure Sensitivity (SS) and False Positivity Rate per hour (FPR/h) are measured from the recorded data signals. Seizure Sensitivity (SS) is the ratio between the (number of predicted seizures)/(total number of seizures) = (number of true alarms)/(total number of seizures). FPR/h is the number of alarms that do not correspond to seizures raised in one hour. FPR/h = ((Number of false alarms/Interictal Duration) - (Number of False Alarms × Refractory period)).
Secondary outcomes
Qualitative Satisfaction Survey
At the end of the study, patients' experience and perception of the EarSD device are collected using a paper-based 7-question survey measured on a 5-point Likert scale ranging from Strongly Disagree to Strongly Agree. A maximum total point score of 35 represents a better reported satisfactory score from participants and having a good experience with the device and its comfortability for daily activities. The survey is a self-administered report, and participants will be asked about the comfortability and perceived utility of the device.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Felicia Chu
Lead sponsor
University of Massachusetts, Worcester
Sponsor institution
University of Massachusetts, Amherst