About this trial
This prospective observational study aims to determine whether a pressure signal recorded through a standard esophageal temperature probe can track changes in cardiac output during general anesthesia.
During surgery, anesthesiologists need to keep the heart adequately filled (preload) to maintain blood flow to vital organs. Central venous pressure, the traditional bedside measure, reflects intravascular volume poorly. Pulmonary capillary wedge pressure, which estimates left atrial pressure, requires a pulmonary artery catheter, an invasive device that is now rarely used. Dynamic indices such as stroke volume variation and pulse pressure variation are reliable only under strict conditions, such as controlled ventilation with an adequate tidal volume, absence of spontaneous breathing, and a regular heart rhythm.
The esophagus lies directly behind the left atrium. When an esophageal probe is filled with saline and connected to a pressure transducer, small pulsatile "cardiac oscillation" waves can be recorded at a certain depth. The investigators hypothesize that this signal reflects pressure changes transmitted from the left atrium and therefore carries real-time information about cardiac filling and output.
Forty adult patients (aged 20 to 79 years, ASA physical status I to III) scheduled for surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring will be enrolled after written informed consent. After induction of anesthesia and tracheal intubation, an 18 Fr esophageal temperature probe, which is routinely used for temperature monitoring, will be filled with 7 to 8 mL of sterile saline, connected to a pressure transducer, and advanced until a clear and stable cardiac oscillation waveform with an amplitude of at least 0.5 to 1.0 mmHg is identified. Anesthesia and surgical care will follow standard practice, and the study will not change clinical management. Arterial pressure, heart rate, esophageal pressure, cardiac output, cardiac index, stroke volume, stroke volume index, stroke volume variation, pulse pressure variation, airway pressure, and positive end-expiratory pressure will be recorded continuously and analyzed around hemodynamic events such as fluid administration, bleeding, and vascular clamping.
The primary outcome is the concordance rate between the percentage changes in stroke volume index and in cardiac oscillation-derived esophageal pressure during these events, assessed by four-quadrant plot analysis after excluding small changes (less than 15%). Secondary outcomes are trending ability assessed by polar plot analysis (angular error), agreement between the percentage changes of the two measures assessed by Bland-Altman analysis (percentage error), and the ability of esophageal pressure changes to predict fluid responsiveness (an increase in stroke volume of more than 15%) assessed by receiver operating characteristic curve analysis.
If the esophageal pressure signal reliably tracks changes in cardiac output, it could offer a simple hemodynamic monitoring method that uses equipment already in routine use during general anesthesia, without additional invasive devices.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Adult patients scheduled for surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring
Age 20 years or older and younger than 80 years
American Society of Anesthesiologists (ASA) physical status I to III
Voluntary written informed consent to participate in the study
Disqualifiers
Esophageal disease, such as esophageal varices, esophageal diverticulum, esophageal stricture, or esophageal cancer
Anatomical variation in the relative position of the esophagus and heart, such as severe chest wall deformity
Significant cardiac functional or structural abnormality on echocardiography, including left ventricular ejection fraction less than 40%, moderate or greater mitral or aortic valve disease, or intracardiac shunt
Persistent atrial fibrillation or frequent premature ventricular contractions
Trial population
Adult patients aged 20 to 79 years with ASA physical status I to III who are scheduled for surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring as part of routine clinical care at Asan Medical Center, Seoul, Republic of Korea. Eligible patients will be approached consecutively before surgery, and those who provide written informed consent will be enrolled.
Trial design
Cohort
Prospective
Treatments tested in this trial
Not listed
Trial groups
Trial outcomes
Primary outcomes
Concordance rate between changes in stroke volume index and cardiac oscillation-derived esophageal pressure
For each intraoperative hemodynamic event (e.g., fluid administration, bleeding, vascular clamping), the percentage change from the pre-event value to the post-event value is calculated for stroke volume index (SVI), measured by arterial pressure-based cardiac output monitoring, and for cardiac oscillation-derived esophageal pressure, measured through a saline-filled esophageal temperature probe. Paired changes are plotted on a four-quadrant plot. After excluding pairs with a change of less than 15% (exclusion zone), the concordance rate is calculated as the percentage of paired changes in which both measures change in the same direction (both increase or both decrease).
Secondary outcomes
Angular error of trending ability assessed by polar plot analysis
Paired percentage changes in stroke volume index and cardiac oscillation-derived esophageal pressure during hemodynamic events are converted into polar coordinates. The mean angular bias (degrees) and the radial limits of agreement (degrees) are calculated to quantify the trending ability of esophageal pressure relative to stroke volume index. Pairs within the 15% exclusion zone are excluded.
Percentage error between changes in stroke volume index and esophageal pressure assessed by Bland-Altman analysis
Because stroke volume index and esophageal pressure have different units, agreement is assessed using the percentage changes of each measure during hemodynamic events. Bland-Altman analysis is performed to calculate the bias and 95% limits of agreement between the paired percentage changes, and the percentage error is derived from these values.
Area under the receiver operating characteristic curve (AUC) of esophageal pressure change for predicting fluid responsiveness
Fluid responsiveness is defined as an increase in stroke volume of more than 15% after fluid administration. The ability of the percentage change in cardiac oscillation-derived esophageal pressure to discriminate fluid responders from non-responders is assessed by receiver operating characteristic curve analysis. The AUC with its 95% confidence interval and the optimal cut-off value (Youden index) are reported.