About this trial
Occlusion pressure at 100 ms (P0.1) is a reliable and non-invasive index of inspiratory drive that can be measured easily on most ventilator. Unfortunately, P0.1 cannot be measured reliably in non-intubated patients. In those receiving non-invasive ventilation, the measurement of P0.1 might be inaccurate due to leaks. P0.1 cannot be measured at all in patients receiving high-flow nasal canulae (HFNC).
Ultrasound indices of diaphragm contractile function, namely diaphragm excursion, velocity and thickening fraction, could be reliable and non-invasive proxies of respiratory drive in non-intubated patient.
Our hypothesis is that ultrasound indices of diaphragm function may reliably estimate P0.1. To validate this hypothesis, theses indices will be measured simultaneously with P0.1 in intubated patients, during diaphragm loading conditions similar to spontaneous breathing: the spontaneous breathing trial (SBT). The SBT is used to evaluate if patients can breathe without the assistance provided by the ventilator, and thus, be weaned from mechanical ventilation.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
- Age ≥ 18 years old
Patient intubated or tracheostomized
Decision by the clinician in charge to initiate a SBT
After information, the patient or next of kind did not refuse to participate (according to the French law, written informed consent is waived)
Disqualifiers
- Suspicion of diaphragmatic nerves injury
Preexisting neuromuscular disease
Spinal cord injury above C5
Neuromuscular blockade within last 24 hours
Trial population
Patients admitted to the intensive care unit who are intubated or have undergone a tracheostomy.
Trial design
Cohort
Prospective
Treatments tested in this trial
Spontaneous breathing trial
DeviceA plateau pressure measurement will be performed before starting the spontaneous breathing trial (SBT). Then, immediately and 5, 10, 15, 20, 25, and 30 minutes after the start of the SBT, the following will be performed as part of the study: * Subcostal diaphragmatic ultrasound: M-mode and tissue Doppler * Lateral diaphragmatic ultrasound: 2D mode and M-mode * Measurement of P0.1 * Measurement of Pocc during a tele-expiratory pause, only 5 minutes after the start of the SBT.
Treatment groups
Trial outcomes
Primary outcomes
P0.1 and Peak contraction velocity 5 minutes after the initiation of the SBT
Airway occlusion pressure measured 100 ms after the initiation of the inspiration, compared to peak contraction velocity of the diaphragm, measured by subcostal ultrasound in TDI-mode
Secondary outcomes
P0.1 and EXdi
Airway occlusion pressure measured 100 ms after the initiation of the inspiration, compared to maximal excursion of the diaphragm, measured by subcostal ultrasound in M-mode
P0.1 and Peak contraction velocity
Airway occlusion pressure measured 100 ms after the initiation of the inspiration, compared to peak contraction velocity of the diaphragm, measured by subcostal ultrasound in TDI-mode
P0.1 and EXdi,0.1
Airway occlusion pressure measured 100 ms after the initiation of the inspiration, compared to excursion of the diaphragm measured 100 ms after the beginning of the inspiration, measured by subcostal ultrasound in M-mode
P0.1 and Velocity Time Integral
Airway occlusion pressure measured 100 ms after the initiation of the inspiration, compared to velocity time integral of the diaphragm, measured by subcostal ultrasound in TDI-mode