About this trial
The goal of this clinical trial is to learn if the Gixam tongue-imaging test gives the same result each time it is done, in adults aged 45 to 84 who are at average risk for colorectal cancer and are scheduled for a screening colonoscopy. The main questions it aims to answer are:
* Does Gixam give the same result when the test is repeated on the same day, with the same device and the same staff member (repeatability)? * Does Gixam give the same result when the test is done on different days, with different devices and different staff members (reproducibility)?
Participants will:
* Answer questions about their health, medicines, diet, alcohol and tobacco use, and oral care habits * Have the Gixam tongue test 6 times over 2 weeks: 4 times on Day 1, once on Day 7 and once on Day 14 * On Days 7 and 14, report when they last ate, drank or smoked, and whether they brushed, scraped their tongue or used mouthwash * Then join the main Gixam study (CL-1009) for a FIT stool test and their planned colonoscopy
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Participants aged ≥45 - ≤84 years.
Able to provide a signed informed consent.
Considered by a physician or healthcare provider as being of 'average risk' for CRC.
Scheduled for a screening colonoscopy at investigation site.
Disqualifiers
Undergoing colonoscopy for investigation of symptoms.
Has undergone colonoscopy within preceding nine (9) years except for a failed colonoscopy due to poor bowel preparation. Failed colonoscopy must have been within the past year and without therapeutic intervention.
Positive FIT or Fecal Occult Blood Test (FOBT) result within the preceding eleven (11) months.
Positive Cologuard, Sheild, ColoSense or Epi proColon test within the preceding three (3) years.
Trial design
Single group
Treatments tested in this trial
Gixam System
DeviceRepeated Gixam tests, 6 times over 2 weeks: 4 times on Day 1, once on Day 7 and once on Day 14
Treatment groups
Trial outcomes
Primary outcomes
Primary Safety Endpoint
Number of device-related adverse events and serious adverse events.
Gixam repeatability - precision within-system-within-operator of all Runs
Success criteria 1. NPA lower limit of the two-sided 95% confidence interval \> 85% 2. PPA lower limit of the two-sided 95% confidence interval \> 80%
Gixam reproducibility - precision between-system-between-operator of all Systems/Operators
Success criteria 1. NPA lower limit of the two-sided 95% confidence interval \> 80% 2. PPA lower limit of the two-sided 95% confidence interval \> 75%
Secondary outcomes
Gixam reported continuous probability score
This endpoint will be used to calculate variability across the repeated assessments (standard deviations and coefficients of variation).