Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorUniversity Hospital, Montpellier

About this trial

SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.

Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.

The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.

This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Adult subject

Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined

Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.

Disqualifiers

Subject lacking the cognitive capacity to perform and understand the experimental protocol

Subject unable to read and/or write

Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used

Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb

Trial design

Design model

Single group

Treatments tested in this trial

  • K-Move/K-Push assessment system

    Device

    Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility.

Treatment groups

40 Participants
are divided into 1 treatment group
Group A: Instrumented spasticity assessmentExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment

Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2.

Time frame
Day 1 and Day 2

Secondary outcomes

1

Difference in total elbow range of motion between the affected and unaffected upper limbs

Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Time frame
Day 1
2

Difference in the mean stretch velocity between the affected and unaffected upper limbs

Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Time frame
Day 1
3

Difference in the maximum resistance force between the affected and unaffected upper limbs

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Time frame
Day 1
4

Difference in the mean resistance force between the affected and unaffected upper limbs

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Time frame
Day 1

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

This trial is not recruiting at the moment. You can still explore other options: