About this trial
Coronary stent implantation is a common treatment for coronary artery disease. However, current methods used to evaluate implanted stents and detect complications such as in-stent restenosis are generally invasive and may require ionizing radiation.
The WAEVE® device is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to assess implanted coronary stents. The NIMBLE-II study is designed to evaluate the performance of a new pre-measurement feature incorporated into WAEVE® that detects heartbeat-synchronous microwave reflections and helps distinguish true stent-related signals from non-cardiac or artifactual signals. By confirming the cardiac origin of the detected reflection, this feature is intended to improve the robustness and reliability of WAEVE® measurements and reduce the risk of false detections arising from non-cardiac or artifactual reflections.
This prospective, multicenter, single-arm clinical investigation will enroll adults undergoing elective percutaneous coronary intervention (PCI). Participants will undergo WAEVE® measurements before PCI and, when a single coronary stent is implanted, after PCI. The study aims to generate technical and clinical evidence supporting the reliability and robustness of the WAEVE® pre-measurement feature under real clinical conditions.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Adults ≥18 years of age.
scheduled for a PCI procedure with potential stent implantation and without any previously implanted coronary stent.
Clinically stable condition allowing completion of pre-PCI WAEVE® measurements.
Ability to understand the study procedures and provide written informed consent.
Disqualifiers
Hemodynamic instability, cardiogenic shock, or severe arrhythmias.
Active implantable electronic devices or metallic implants on the chest that may interfere with measurements (e.g., pacemakers, defibrillators, CRT devices, sternal wires, rib plates, metallic clips).
Presence of skin conditions, wounds, or lesions in the areas designated for transducer placement (including extensive thoracic skin lesions).
Inability to remain supine or still during WAEVE® measurements.
Trial design
Single group
Treatments tested in this trial
WAEVE®
DeviceWAEVE® is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to detect and analyze microwave reflections associated with implanted coronary stents. Participants undergo WAEVE® measurements before PCI and, when applicable, after coronary stent implantation. The study evaluates the performance of a pre-measurement feature designed to detect heartbeat-synchronous microwave reflections and distinguish true cardiac and stent-related signals from non-cardiac or artifactual reflections.
Treatment groups
Trial outcomes
Primary outcomes
Heartbeat-Synchronous Signal Detection Rate in Post-PCI Acquisitions
Proportion of assessable post-PCI pre-measurement acquisitions in which a heartbeat-synchronous microwave signal is detected, defined by the presence of statistically significant periodic modulation of the microwave signal at the cardiac frequency using the electrocardiogram (ECG) signal as the temporal cardiac-cycle reference.
Secondary outcomes
False Detection Rate
Proportion of assessable acquisitions exhibiting stent-compatible microwave patterns that do not demonstrate heartbeat-linked periodicity and are therefore classified as non-cardiac or artifactual reflections.
Intra-Patient Measurement Repeatability
Repeatability of repeated WAEVE acquisitions performed under identical acquisition conditions within the same session, quantified using the Measurement Stability Index (MSI). The Measurement Stability Index ranges from 0 to 1, with higher values indicating greater agreement between repeated acquisitions
Proportion of Technically Assessable Acquisitions
Proportion of acquisitions meeting predefined signal-quality criteria and considered suitable for analysis
Signal Changes Associated With Stent Implantation
Within-subject differences between pre-PCI and post-PCI microwave reflection characteristics at matched anatomical sites, summarized using reflection morphology features and the Peak Prominence Difference Index (PPDI).
Other outcomes
Association Between Inflammatory Biomarkers and WAEVE® Signal Parameters
Exploratory correlations between inflammatory biomarker levels and selected WAEVE® signal characteristics, including heartbeat-synchronous signal metrics and stent-related reflection features.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Nimble Diagnostics S.L.
Lead sponsor
Fundación EPIC
Collaborator