Evaluation of WAEVE® Pre-Measurement Feature for Detection of Implanted Coronary Stents in Patients Undergoing Elective PCI

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorNimble Diagnostics S.L.

About this trial

Coronary stent implantation is a common treatment for coronary artery disease. However, current methods used to evaluate implanted stents and detect complications such as in-stent restenosis are generally invasive and may require ionizing radiation.

The WAEVE® device is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to assess implanted coronary stents. The NIMBLE-II study is designed to evaluate the performance of a new pre-measurement feature incorporated into WAEVE® that detects heartbeat-synchronous microwave reflections and helps distinguish true stent-related signals from non-cardiac or artifactual signals. By confirming the cardiac origin of the detected reflection, this feature is intended to improve the robustness and reliability of WAEVE® measurements and reduce the risk of false detections arising from non-cardiac or artifactual reflections.

This prospective, multicenter, single-arm clinical investigation will enroll adults undergoing elective percutaneous coronary intervention (PCI). Participants will undergo WAEVE® measurements before PCI and, when a single coronary stent is implanted, after PCI. The study aims to generate technical and clinical evidence supporting the reliability and robustness of the WAEVE® pre-measurement feature under real clinical conditions.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Adults ≥18 years of age.

scheduled for a PCI procedure with potential stent implantation and without any previously implanted coronary stent.

Clinically stable condition allowing completion of pre-PCI WAEVE® measurements.

Ability to understand the study procedures and provide written informed consent.

Disqualifiers

Hemodynamic instability, cardiogenic shock, or severe arrhythmias.

Active implantable electronic devices or metallic implants on the chest that may interfere with measurements (e.g., pacemakers, defibrillators, CRT devices, sternal wires, rib plates, metallic clips).

Presence of skin conditions, wounds, or lesions in the areas designated for transducer placement (including extensive thoracic skin lesions).

Inability to remain supine or still during WAEVE® measurements.

Trial design

Design model

Single group

Treatments tested in this trial

  • WAEVE®

    Device

    WAEVE® is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to detect and analyze microwave reflections associated with implanted coronary stents. Participants undergo WAEVE® measurements before PCI and, when applicable, after coronary stent implantation. The study evaluates the performance of a pre-measurement feature designed to detect heartbeat-synchronous microwave reflections and distinguish true cardiac and stent-related signals from non-cardiac or artifactual reflections.

Treatment groups

120 Participants
are divided into 1 treatment group
Group A: WAEVE® AssessmentExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Heartbeat-Synchronous Signal Detection Rate in Post-PCI Acquisitions

Proportion of assessable post-PCI pre-measurement acquisitions in which a heartbeat-synchronous microwave signal is detected, defined by the presence of statistically significant periodic modulation of the microwave signal at the cardiac frequency using the electrocardiogram (ECG) signal as the temporal cardiac-cycle reference.

Time frame
Day 1 (day of the PCI procedure)

Secondary outcomes

1

False Detection Rate

Proportion of assessable acquisitions exhibiting stent-compatible microwave patterns that do not demonstrate heartbeat-linked periodicity and are therefore classified as non-cardiac or artifactual reflections.

Time frame
Day 1 (day of the PCI procedure)
2

Intra-Patient Measurement Repeatability

Repeatability of repeated WAEVE acquisitions performed under identical acquisition conditions within the same session, quantified using the Measurement Stability Index (MSI). The Measurement Stability Index ranges from 0 to 1, with higher values indicating greater agreement between repeated acquisitions

Time frame
Day 1 (day of the PCI procedure)
3

Proportion of Technically Assessable Acquisitions

Proportion of acquisitions meeting predefined signal-quality criteria and considered suitable for analysis

Time frame
Day 1 (day of the PCI procedure)
4

Signal Changes Associated With Stent Implantation

Within-subject differences between pre-PCI and post-PCI microwave reflection characteristics at matched anatomical sites, summarized using reflection morphology features and the Peak Prominence Difference Index (PPDI).

Time frame
Day 1 (day of the PCI procedure)

Other outcomes

1

Association Between Inflammatory Biomarkers and WAEVE® Signal Parameters

Exploratory correlations between inflammatory biomarker levels and selected WAEVE® signal characteristics, including heartbeat-synchronous signal metrics and stent-related reflection features.

Time frame
Day 1 (day of the PCI procedure)

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Nimble Diagnostics S.L.

Lead sponsor

Fundación EPIC

Collaborator