About this trial
This study aims to evaluate the feasibility and diagnostic accuracy of the CART-BP ring-type wearable blood pressure monitoring device in comparison with 24-hour ambulatory blood pressure monitoring (ABPM) in patients with heart failure with reduced ejection fraction (HFrEF, LVEF ≤ 40%). As a multi-center, prospective, exploratory study, 100 patients will be enrolled at two tertiary hospitals in South Korea. The agreement between the two devices in 24-hour mean, daytime, and nighttime blood pressure measurements will be assessed per ISO 81060-2:2018 criteria.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Age ≥ 19 years (male or female)
Diagnosed with heart failure and currently under treatment
Left ventricular ejection fraction (LVEF) ≤ 40% confirmed by echocardiography performed within 1 year prior to enrollment
Able to undergo both 24-hour ambulatory blood pressure monitoring and ring-type blood pressure measurement
Disqualifiers
Office blood pressure: SBP < 90 mmHg, or DBP < 50 mmHg, or SBP > 180 mmHg, or DBP > 100 mmHg
Unable to perform 24-hour ambulatory blood pressure monitoring
Unable to wear or use the ring-type blood pressure device
Variability in three consecutive resting cuff blood pressure measurements ≥ 20 mmHg (SBP) or ≥ 10 mmHg (DBP)
Trial population
Heart Failure with Reduced Ejection Fraction patients
Trial design
Cohort
Prospective
Treatments tested in this trial
Not listed
Trial groups
Trial outcomes
Primary outcomes
Agreement between CART-BP and ABPM in 24-hour mean systolic and diastolic blood pressure
Agreement between the CART-BP ring-type device and the 24-hour ABPM for mean 24-hour systolic blood pressure (SBP) and diastolic blood pressure (DBP), evaluated according to ISO 81060-2:2018 Criteria 1 and 2 (Opposite arm simultaneous method, section 5.2.4.2.2)
Secondary outcomes
Agreement in daytime mean SBP and DBP between CART-BP and ABPM
Agreement in nighttime mean SBP and DBP between CART-BP and ABPM
Agreement in blood pressure dipping pattern between CART-BP and ABPM
Blood pressure variability (standard deviation) agreement between devices