About this trial

Rapid On-Site Evaluation (ROSE) of cytological slides obtained during EBUS-TBNA improves diagnostic yield by providing real-time adequacy assessment and preliminary diagnostic orientation after each needle pass. In centers without a dedicated cytopathologist, ROSE is performed by a second interventional pulmonologist acting as a dedicated ROSE operator (interventional pulmonologist-performed ROSE, IP-ROSE), a model associated with good but variable diagnostic performance compared to cytopathologist-performed ROSE.

This study evaluates the feasibility and preliminary diagnostic performance of an investigational artificial intelligence prototype for digital ROSE. The prototype, developed in-house by the Principal Investigator, analyzes microscopic images of Diff-Quik stained cytological slides acquired through a dedicated digital microscope, together with basic clinical data, via API calls to a multimodal AI model. It produces two outputs: sample adequacy (appropriate/not appropriate) and malignancy suspicion (benign/malignant), each with a confidence score. The AI output is recorded in the study database for research purposes only and is not shown to the operator in real time; it does not influence clinical decisions during the procedure.

The study is a prospective, monocentric, observational pilot study enrolling 65 adult patients undergoing EBUS-TBNA or peripheral TBNA with IP-ROSE at a single interventional pulmonology unit. The primary statistical unit is the individual ROSE slide, with an expected 130 to 160 evaluable slides. Co-primary endpoints are: (1) technical feasibility of the AI prototype, defined as the proportion of slides with valid AI output within 90 seconds; and (2) AI accuracy for sample adequacy assessment compared to the definitive cytopathological diagnosis, with an expected 95% confidence interval precision of ±5.5%. Secondary endpoints include AI accuracy for malignancy suspicion, agreement between the AI prototype and the IP-ROSE operator, and AI output latency.

The AI prototype is not a commercially approved or CE-marked medical device. It was developed internally by the Principal Investigator for research purposes and is evaluated exclusively within this study. Data from this pilot study will inform the design of a subsequent confirmatory non-inferiority trial, which will be the subject of separate registration and ethical approval.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Age ≥ 18 years.

Clinical indication for diagnostic bronchoscopy with EBUS-TBNA for mediastinal lymphadenopathy and/or TBNA on peripheral lung lesions, according to ACCP/ERS-ESTS guidelines and the Unit's diagnostic pathways.

The procedure will be performed in two-person mode with IP-ROSE performed as per the center's SOP.

Ability to provide written informed consent.

Disqualifiers

Absolute contraindications to bronchoscopy.

Known histological diagnosis of the target lesion (except re-staging).

Concomitant interventional procedures that alter the standard sequence.

Inability to provide written informed consent.

Trial population

patients with clinical indication for diagnostic bronchoscopy with EBUS-TBNA for mediastinal lymphadenopathy and/or TBNA on peripheral lung lesions, according to ACCP/ERS-ESTS guidelines and the Unit's diagnostic pathways.

Trial design

Design model

Case-only

Time perspective

Prospective

Treatments tested in this trial

  • ROSE

    Diagnostic test

    Bronchologist's Rapid On Site Evaluation (ROSE) compared with those of the AI and then compared with the pathological anatomy report

Treatment groups

65 Participants
are divided into 1 treatment group
Group A: patients with an indication for diagnostic bronchoscopy1 intervention

Trial outcomes

Primary outcomes

1

technical feasibility of the AI prototype

technical feasibility of the AI prototype, defined as the proportion of slides with valid AI output within 90 seconds

Time frame
6 months
2

AI accuracy

AI accuracy for sample adequacy assessment compared to the definitive cytopathological diagnosis, with an expected 95% confidence interval precision of ±5.5%

Time frame
6 months

Secondary outcomes

1

AI accuracy for malignancy suspicion

AI accuracy for malignancy suspicion

Time frame
6 months
2

Agreement between the AI prototype and the IP-ROSE operator

Agreement between the AI prototype and the IP-ROSE operator

Time frame
6 months
3

AI output latency

AI output latency

Time frame
6 months

Other outcomes