Feasibility of End-tidal Oxygen Concentration Monitoring During Preoxygenation for Intubation in the Intensive Care Unit.

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorCentre Hospitalier Régional d'Orléans

About this trial

Intubation is a common procedure in the intensive care unit. Hypoxemia is the most frequent complication of this procedure.

Monitoring the end-tidal of oxygen is recommended in operating room (OR). End-tidal of oxygen (EtO2) \>90% is an indication of a correct preoxygenation.

This monitoring is not used in routine in intensive care unit (ICU). There is no recommendation on the monitoring of end-tidal of oxygen in intensive care unit. In practice, clinicians use pulsed oxygen saturation (SpO2) to determine whether the patient is sufficiently preoxygenated. However, this parameter is not a good indicator of a correct preoxygenation.

In the OR, patients are compliant during the preoxygenation period and the measure of EtO2 with the face mask monitor is considered reliable because i) mask leakage is minimal and ii) the patient can breathe slowly and regularly.

Theses conditions are not available in critical ill patients requiring emergency intubation. EtO2 measured on the facemask may not reflect true EtO2.

This concern about the reliability of EtO2 measurement via the facemask justifies that we conduct a study to compare EtO2 measured on a facemask (facemask EtO2) to EtO2 measured in pharynx (via e nasopharyngeal catheter).

The aim of this study is to determine whether the measurement of EtO2 on facemask is reliable in patients in ICU.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Age ≥ 18 years

Requiring an intubation in ICU for acute respiratory failure defined as a respiratory rate ≥ 25/min or clinical signs of respiratory distress or hypoxemia defined as oxygen requirement ≥ 15L/min or FiO2 ≥80% to maintain SpO2 ≥92% or a PaO2/FiO2 ratio <100 mmHg

With preoxygenation by noninvasive ventilation

Disqualifiers

Intubation for cardiac arrest

EtO2 monitoring not available

Pulsed oxygen saturation monitoring specific for the study not available

Preoxygenation by high-flow nasal oxygen therapy or bag-mask ventilation

Trial design

Design model

Single group

Treatments tested in this trial

  • double monitoring of EtO2

    Other intervention

    A double monitoring of EtO2 will be performed during the procedure of intubation (from the beginning of the preoxygenation to the success of intubation): * In the pharynx (via a nasopharyngeal catheter) for the needs of the study * On the facemask (as the practice in the OR)

Treatment groups

30 Participants
are divided into 1 treatment group
Group A: monitoring of the expired oxygen fractionExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Percentage of expired O2 at facemask during preoxygenation

compare facemask EtO2 to pharyngeal EtO2 during preoxygenation in order to know if facemask monitoring is sufficient or if a nasopharyngeal catheter is absolutely necessary to monitor alveolar oxygenation.

Time frame
Hour 1
2

Percentage of expired pharyngeal O2 during preoxygenation

compare facemask EtO2 to pharyngeal EtO2 during preoxygenation in order to know if facemask monitoring is sufficient or if a nasopharyngeal catheter is absolutely necessary to monitor alveolar oxygenation.

Time frame
Hour 1

Secondary outcomes

1

Number of patient who had a pulsed oxygen saturation under 90%

Hypoxemia is defined as the occurrence of a pulsed oxygen saturation (SpO2) ≤ 90%. The end of preoxygenation is defined by the induction of anesthetics drugs. Successful intubation is defined by the presence of 3 identical waveforms on the capnograph.

Time frame
Hour 1

Other outcomes