About this trial
The goal of this clinical trial is to assess if the TRICENTO G2 Transcatheter Valve System for Tricuspid Regurgitation (TRICENTO G2 TVSTR) is safe and efficient in treating severe tricuspid regurgitation in adult patients.
The main questions it aims to answer are:
* Does the treatment with the TRICENTO G2 TVSTR improve symptoms of severe tricuspid regurgitation? * Is the treatment with the TRICENTO G2 TVSTR safe?
Participants will:
* Undergo a minimally invasive procedure for the implantation of the TRICENTO G2 Bioprosthesis using the TRICENTO G2 Delivery System * Visit the clinic for a baseline visit, the procedure itself, 30 days, 3 months, 6 months and 1 year after the procedure
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Patients with symptomatic tricuspid regurgitation (TR), graded as severe or greater (accord. to TR 5-Tier Grading System TVARC)
Patients with TR leading to NYHA class II, III or IV or clinical signs of right heart failure
Patients suitable for implantation of the TRICENTO G2 Bioprosthesis according to anatomical and clinical criteria assessed by computed tomography, echocardiography and/or right heart catheterization (confirmed by a heart team)
Patients receiving optimal medical treatment for underlying diseases
Disqualifiers
Superior vena cava diameter < 22.9 mm or > 37.1 mm (perimeter derived)
Inferior vena cava diameter < 27.9 mm or > 42.9 mm (perimeter derived)
Right atrium cavity height < 55 mm or > 80 mm (direct path between caval ostia)
Peak right atrial pressure > 50 mmHg
Trial design
Single group
Treatments tested in this trial
TRICENTOG2 Transcatheter Valve System for Tricuspid Regurgitation (TRICENTOG2 TVSTR)
DeviceThe TRICENTO G2 TVSTR consists of a Bioprosthesis and a Delivery System including a Loading Funnel. The Bioprosthesis is an endovascular graft-like implant in combination with a lateral valve element. It is placed into the right atrium while spanning from inferior to superior vena cava leaving the native valve apparatus untouched (heterotopic approach). The TRICENTO G2 Bioprosthesis is designed to reduce the reflux of blood into the caval system during systole while allowing for forward flow during diastole. The TRICENTO G2 Bioprosthesis is delivered minimally invasive by transfemoral venous access using a catheter-based Delivery System.
Treatment groups
Trial outcomes
Primary outcomes
Technical success: Percentage of participants with successful access, delivery and retrieval of the TRICENTO G2 Delivery System AND successful deployment and correct positioning of the TRICENTO G2 Bioprosthesis measured via echocardiography
Technical success \[given as percentage of participants\] 1. Successful access, tracking and withdrawal of the TRICENTO G2 Delivery System \[Yes/No\] 2. Successful deployment and correct positioning of the TRICENTO G2 Bioprosthesis \[Yes/No\]
Frequency and composite of device-related or procedure-related serious adverse events (SAEs)
With SAE classification according to MDR and MDCG 2020-10.
Frequency of device deficiencies
Device deficiency: Inadequacy of a medical device with respect to its identity, quality, durability, reliability, usability, safety or performance.
Rate of unplanned surgery or re-intervention due to a life-threatening device or procedure failure
Secondary outcomes
Clinical performance and functionality of the TRICENTO G2 Bioprothesis assessed via echocardiography
Assessment via echocardiography (TEE/TTE): * Fully functional bioprosthetic valve demonstrating unrestricted inflow in diastole and no backflow in systole beyond the closing volume * Exclusion of para-prosthetic leakage * Changes in hepatic vein systolic backflow
Clinical performance and functionality of the TRICENTO G2 Bioprothesis assessed via right heart catheterization
Change in peak central venous pressure (CVP \[mmHg\]) or CVP pulsatility or differential pressure \[mmHg\] over the prosthetic valve right atrium-central venous system.
Clinical performance and functionality: Assessment of device position
Assessment of device position assessed via computed tomography (CT).
Clinical performance and functionality: Assessment of stent geometry
Assessment of stent geometry assessed via computed tomography (CT).
Other outcomes
Exploartory safety endpoint: Overall mortality
All deaths occurring during the clinical investigation.
Exploratory performance endpoint: Changes in Tricuspid Regurgitation - Effective Regurgitant Orifice Area (TR EROA)
Imaging of cardiac structures via echocardiography, comparing TR EROA \[mm\^2\] to baseline.
Exploratory performance endpoint: Changes in TR regurgitant volume
Imaging of cardiac structures via echocardiography, comparing TR regurgitant volume \[mL\] to baseline.
Exploratory performance endpoint: Changes in Tricuspid annular plane systolic excursion (TAPSE)
Imaging of cardiac structures via echocardiography, comparing TAPSE \[mm\] to baseline.