Handheld Point of Care Electro-Diagnostic Device

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorIndiana University

About this trial

There are currently several products offering POC EEG, intended to be deployed on each patient individually and continuously. The investigators propose studying a different approach by using a migrating POC electrodiagnostic device that is deployed directly by the onsite provider and is actively used, like other POC devices, to acquire real-time diagnostic insight and facilitate immediate medical decision making. This can significantly shorten time to treatment which in several neurological and neurosurgical conditions, and thus, can significantly improve patient outcome.

The investigational POC investigational device is based on a handheld limited electrode array, complemented by various stimulation modalities such as SSEP, VEP and BAEP, and potentially other sensors. For this study, the investigators intend to use readily available industry standard stimulators and only use the proposed POC investigational device to acquire and analyze the data. The POC investigational electrodiagnostic prototype was assembled using OpenBCI8 electronic board and electrodes, based on ADS 129X chip by Texas Instruments, that was shown to provide an adequate low-cost solution for electrodiagnostic data acquisition.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Patients 18 years old or above

Seen inpatient or in outpatient service at IU Health facilities - Methodist Hospital, and the Neuroscience Center, by a Neurology and/or Neurosurgery consultant as part of their SOC.

The participant has capacity to sign consent or has an assigned LAR (Legally Authorized Representative - Spouse, parent, sibling, etc.) able to consent on their behalf.

Disqualifiers

Pregnant woman

PI unblinded by participant or LAR regarding indication or clinical data

Burns or extensive skin lesions

Implanted stimulation devices (any)

Trial design

Design model

Single group

Treatments tested in this trial

  • POC EEG

    Diagnostic test

    The PI will apply the POC investigational device dry electrodes, as well as SOC SSEP stimulator to the participant and run a standardized diagnostic testing sequence, to cover all 4 limbs, spine, and head. The device connects to a laptop via Bluetooth, and the acquired electrodiagnostic data is visualized on the screen. The PI will analyze the acquired data and determine lateralization, localization, and presumed nature of the lesion or pathology causing the patient's symptoms. The direct contact with the participant should take10-20 minutes. The inferred lateralization, localization, and nature of the presumed lesion, will be then documented, to be compared to the diagnostic insight of using SOC diagnostic methods such as CT, MRI, EEG, SSEP, etc. The participant will receive their regular SOC management and a single, brief evaluation using the POC investigational device.

Treatment groups

30 Participants
are divided into 1 treatment group
Group A: Treatment ArmExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Outcome 1

The diagnostic insights acquired using the POC investigation device regarding lateralization (left/right), localization (cerebral - frontal, temporal, parieto-occipital; spinal - cervical, thoracal, lumbosacral; or radicular), and presumed nature of lesion (i.e. ischemia, hemorrhage, space-occupying lesion), are significantly comparable to, and congruent with, the SOC diagnostic insight regarding same parameters (lateralization, localization and presumed etiology) acquired using SOC techniques such as MRI, 10-20 (standard ubiquity used electrode montage) EEG systems, SSEP, etc. The PI will be the only POC investigational device operator, and will be blinded to patient examination, symptoms and SOC investigation results. The PI will be blinded to all EMR data. The investigator then plans to compare the clinical insight obtained by using the POC investigational device, to SOC.

Time frame
Day 0
2

Outcome 2

The diagnostic insights acquired using the POC investigation device regarding localization (cerebral - frontal, temporal, parieto-occipital; spinal - cervical, thoracal, lumbosacral; or radicular),are significantly comparable to, and congruent with, the SOC diagnostic insight regarding same parameters (lateralization, localization and presumed etiology) acquired using SOC techniques such as MRI, 10-20 (standard ubiquity used electrode montage) EEG systems, SSEP, etc.

Time frame
Day 0
3

Outcome 3

The diagnostic insights acquired using the POC investigation device regarding the presumed nature of lesion (i.e. ischemia, hemorrhage, space-occupying lesion), are significantly comparable to, and congruent with, the SOC diagnostic insight regarding same parameters (lateralization, localization and presumed etiology) acquired using SOC techniques such as MRI, 10-20 (standard ubiquity used electrode montage) EEG systems, SSEP, etc.

Time frame
Day 0

Secondary outcomes

Other outcomes

Sponsors and contacts