HUD Surgical Guidance for Toric Alignment

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorGainesville Eye Associates

About this trial

This study aims to assess the accuracy of toric intraocular lens (IOL) alignment axes as calculated and displayed by a novel digital guidance system. In addition, it evaluates the time required to initiate surgical guidance and postoperative refractive astigmatism following toric IOL implantation.

This single-surgeon, single-site, prospective, non-invasive observational study evaluated the accuracy of toric IOL alignment axes calculated and displayed by the digital guidance system (Cassini Technologies, B.V.) in 100 eyes. The intraoperatively calculated alignment axes were compared to preoperative measurements to assess their accuracy (in degrees). Postoperative outcomes, including residual refractive astigmatism (diopters) and visual acuity (Snellen), were recorded at 1 month postoperatively.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Patients undergoing uncomplicated cataract surgery with toric IOL implantation,

Eyes implanted with a toric IOL.

Disqualifiers

Ocular comorbidity likely to impair postoperative visual acuities, or

History of ocular trauma or zonular instability, or

Previous refractive surgery, or

Irregular corneal astigmatism or keratoconus, or

Trial design

Design model

Single group

Treatments tested in this trial

  • Cassini Guidance

    Device

    Cassini Guidance with HUD

Treatment groups

100 Participants
are divided into 1 treatment group
Group A: Surgical guidance for Toric IOL alignment intraoperativelyOther 1 intervention

Trial outcomes

Primary outcomes

1

Accuracy (measured in degrees of arc) using the Cassini Guidance system for calculated alignment axis for toric IOL implantation.

Primary Endpoint: 1\. Accuracy of the calculated alignment axis (degrees of arc = angular degrees) * Definition: The angular difference between the alignment axis for toric IOLs as determined with the CGS and the intended alignment axis as determined with the Cassini Ambient. * Thresholds of interest: ≤3°, ≤5°, ≤10°, \>10°. * Justification: Misalignment greater than 10° results in a clinically significant loss of toric correction (approximately 33%) (Quesada et al., 2022; Hu, 2023), making this a meaningful accuracy benchmark.

Time frame
Total completion up to 12 months

Secondary outcomes

1

Time efficiency (measured in seconds) of the Cassini Guidance System in eyes undergoing cataract surgery with toric IOL implantation.

Secondary Endpoint: 1\. The time (measured in seconds) required to initiate (0 seconds) and finish (x seconds) surgical guidance using the Cassini Guidance System in 100 eyes undergoing toric IOL implantation.

Time frame
Total completion up to 12 months

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.