In Vivo Treatment Verification of Brachytherapy

ConditionBrachytherapy
Trial statusRecruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
AgeNot listed
SponsorMaastricht Radiation Oncology

About this trial

To acquire data during High Dose Rate (HDR) brachytherapy treatments aiming to track the brachytherapy source inside the patient verifying the delivery of the treatment plan.

Medical images (CT,MRI and US) combined with software developed by the investigator will be employed to predict the response of the Investigational product during the treatment and will be used as reference.The IP will be placed on the side of the treatment table and will acquire data during the whole treatment, which will be compared against information derived from the treatment plan. The IP based method can detect errors and information about motion and treatment uncertainties that are not currently available.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Treatment site Gynaecology (Endometrium/Cervix) and prostate

Treatment: HDR Brachytherapy

Patient will be treated on CT Couch

Mentally competent patient

Disqualifiers

Incapacitated patient

Trial population

The developed method is expected to work for all treatments sites. However, HDR gynecological and prostate treatments are our main interest since they are the most common HDR treatments at Maastro.

Trial design

Design model

Case-only

Time perspective

Prospective

Treatments tested in this trial

  • External imaging panel

    Device

    The IP will be placed on the side of the treatment table and will acquire data during the whole treatment, which will be compared against information derived from the treatment plan. The IP based method can detect errors and information about motion and treatment uncertainties that are not currently available.

Treatment groups

No treatment groups listed

Trial outcomes

Primary outcomes

1

Dwell times

Verify all relevant parameters of the HDR treatment plan during treatment delivery such as dwell times (time in seconds). Brachytherapy treatments are delivered using two parameters, the first parameter is dwell times (the duration the radioactive source stays in a dwell position, measured in seconds (s)). Our system measures dwell times, comparing them against expected values. Note that this parameter can be further used to derive other metrics, such as radiation dose distributions (units in Gray (Gy)), as defined by the TG-43 AAPM dose formalism.

Time frame
1 year
2

Dwell position

Verify all relevant parameters of the HDR treatment plan during treatment delivery such as dwell positions (distance in (milli)meters). Brachytherapy treatments are delivered using two parameters, the second parameter is dwell positions (a location where the radioactive source should stop, defined in centimeters (cm)) Our system measures dwell positions, comparing them against expected values. Note that this parameter can be further used to derive other metrics, such as radiation dose distributions (units in Gray (Gy)), as defined by the TG-43 AAPM dose formalism.

Time frame
1 year

Secondary outcomes

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.