Interstitial Assessment of Architectural Heterogeneity in Prostate Cancer Using a Fine Needle Photoacoustic Probe

Trial statusRecruiting
Trial phaseNot listed
Trial typeObservational
Biological sexMale
Age40-80
SponsorUniversity of Michigan Rogel Cancer Center

About this trial

In this project, we will integrate the Photoacoustic (PA) technology into a prostate biopsy procedure through a fine needle probe. The needle probe will be in the shape of a biopsy needle and compatible with the needle insertion mechanism in the transrectal ultrasound (TRUS) probe. When inserted into the prostate, the fine needle PA probe will assess the histological information in its surrounding tissue without any tissue extraction. The needle probe has been tested in prostate tissue samples and whole human prostates ex vivo.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

1) Planned prostate biopsy as part of routine clinical care

2) > 40 and <80 years old

Disqualifiers

1) Known history of bleeding disorders or

2) Patients taking anticoagulants, no matter doses or washout time

3) Known allergy to silicone material

4) Unable to provide informed consent

Trial population

Patients with planned prostate biopsy procedures will be recruited through U-M Urology Clinics

Trial design

Design model

Other

Time perspective

Prospective

Treatments tested in this trial

  • Fine Needle Photoacoustic Probe

    Device

    Fine Needle Photoacoustic Probe guided by the standard biopsy system without tissue core extraction

Treatment groups

50 Participants
are divided into 1 treatment group
Group A: Biopsy1 intervention

Trial outcomes

Primary outcomes

1

Demonstration of reliable measurements

The primary aim of this study is to demonstrate that our fine needle PA probe can acquire reliable measurements of prostate glands. This will be indicated by a signal-to-noise ratio of the measurements that is larger than one. We will calculate the signal to noise ratio of the signals as the peak-to-peak signal magnitude over the amplitude of the system background signal. The technical failure is that we will not be able to achieve measurements with signal-to-noise ratios larger than 1 in 10% of the measurements. We will have a futility analysis after first 5 patients are completed (thus 10 measurements). If of those 10 measurements, 2 have a SNR less 1, then the study will be stopped as it is unlikely to meet its 90% target.

Time frame
day 1

Secondary outcomes

1

Determining the ability to observe differences between benign and cancerous regions

The secondary aim of this study is documenting our ability to observe differences between the PASA linear slope values derived from benign and cancerous regions, i.e., to detect the difference of 0.05 of 0.19, 0.17, 0.14, 0.20 dB/MHz between the means of the linear slopes in benign and cancerous regions at the wavelengths of 1220 nm, 1370 nm, 800 nm and 266 nm, respectively. We will aim at differentiating the measurements in benign and cancerous regions at specific wavelengths. We have a null hypothesis for each wavelength that at this specific wavelength, our measurement cannot differentiate the benign and cancerous regions. We will test the hypothesis with paired t-tests.

Time frame
day 1