About this trial
In this project, we will integrate the Photoacoustic (PA) technology into a prostate biopsy procedure through a fine needle probe. The needle probe will be in the shape of a biopsy needle and compatible with the needle insertion mechanism in the transrectal ultrasound (TRUS) probe. When inserted into the prostate, the fine needle PA probe will assess the histological information in its surrounding tissue without any tissue extraction. The needle probe has been tested in prostate tissue samples and whole human prostates ex vivo.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
1) Planned prostate biopsy as part of routine clinical care
2) > 40 and <80 years old
Disqualifiers
1) Known history of bleeding disorders or
2) Patients taking anticoagulants, no matter doses or washout time
3) Known allergy to silicone material
4) Unable to provide informed consent
Trial population
Patients with planned prostate biopsy procedures will be recruited through U-M Urology Clinics
Trial design
Other
Prospective
Treatments tested in this trial
Fine Needle Photoacoustic Probe
DeviceFine Needle Photoacoustic Probe guided by the standard biopsy system without tissue core extraction
Treatment groups
Trial outcomes
Primary outcomes
Demonstration of reliable measurements
The primary aim of this study is to demonstrate that our fine needle PA probe can acquire reliable measurements of prostate glands. This will be indicated by a signal-to-noise ratio of the measurements that is larger than one. We will calculate the signal to noise ratio of the signals as the peak-to-peak signal magnitude over the amplitude of the system background signal. The technical failure is that we will not be able to achieve measurements with signal-to-noise ratios larger than 1 in 10% of the measurements. We will have a futility analysis after first 5 patients are completed (thus 10 measurements). If of those 10 measurements, 2 have a SNR less 1, then the study will be stopped as it is unlikely to meet its 90% target.
Secondary outcomes
Determining the ability to observe differences between benign and cancerous regions
The secondary aim of this study is documenting our ability to observe differences between the PASA linear slope values derived from benign and cancerous regions, i.e., to detect the difference of 0.05 of 0.19, 0.17, 0.14, 0.20 dB/MHz between the means of the linear slopes in benign and cancerous regions at the wavelengths of 1220 nm, 1370 nm, 800 nm and 266 nm, respectively. We will aim at differentiating the measurements in benign and cancerous regions at specific wavelengths. We have a null hypothesis for each wavelength that at this specific wavelength, our measurement cannot differentiate the benign and cancerous regions. We will test the hypothesis with paired t-tests.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
University of Michigan Rogel Cancer Center
Lead sponsor
National Cancer Institute (NCI)
Collaborator