About this trial
This study aims to learn more about the information that can be obtained directly from measurements through a Nexa cochlear implant during and after surgery. Researchers will investigate whether these measurements can help understand how the implant is interacting with the inner ear and hearing nerve, where the electrodes are positioned inside the cochlea, and potentially how the implant should be programmed and how well a person detects sounds.
The study will also evaluate how well two methods agree in determining the position of the electrode contacts relative to the anatomy of the inner ear and the hearing nerve. One method uses measurements after surgery made through the implant itself. The other uses medical Computed Tomography(CT) imaging made after the surgery. The information collected will also contribute to a clinical research database that may be used to support future research and development. Participants will be told about this potential future use when they provide consent.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Aged 18 years or older (no upper age limit).
Candidate for CI1032 cochlear implantation as per the local sensorineural hearing loss criteria.
Sufficient proficiency in the language used to conduct the study procedures, as determined by the investigator.
Availability of pre-op CT scan of the cochlear region meeting quality requirements as determined by the sponsor.
Disqualifiers
Previous cochlear implantation in the same ear.
Cochlear implant in the contralateral ear.
Have an unaided air conduction pure tone average (average of unaided thresholds at 0.5, 1, 2 and 4 kHz) ≤40 dB HL in the contralateral ear measured within the last 12 months
Severe or greater sensorineural hearing loss prior to 2 years of age, as documented in medical history.
Trial design
Single group
Treatments tested in this trial
Nexa cochlear implant
DeviceParticipants will be tested with intra- and post-operative objective measures, including Transimpedance Matrix (TIM) and Neural Response telemetry (NRT) recordings, using monopolar and focused multipolar stimulation.
Treatment groups
Trial outcomes
Primary outcomes
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP T-Levels
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative computed tomography (CT) imaging and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP T-levels.
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP C-Levels
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP C-levels.
Correlation Between CT-Derived Electrode-to-Modiolus Distance and Aided Pure-Tone Thresholds
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and aided hearing thresholds.
Correlation Between Monopolar T-NRT Thresholds and MAP T-Levels
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP T-levels.
Secondary outcomes
Mean Absolute Error Between TIM-Derived and CT-Derived Electrode-to-Modiolus Distance
Postoperative electrode-to-modiolus distance for individual electrode contacts will be determined using postoperative computed tomography (CT) imaging and using Transimpedance Matrix (TIM) measurements analysed with an algorithm. The mean absolute error between paired TIM-derived and CT-derived electrode-to-modiolus distance measurements will be reported in millimetres (mm).
This trial is not recruiting at the moment. You can still explore other options: