About this trial
The AGREE-BP study aims to evaluate how closely blood pressure measurements taken with an automated upper-arm cuff agree with measurements taken through a small plastic tube placed in an artery, called an arterial catheter, during the start of general anesthesia.
Blood pressure can change rapidly when patients receive medication to induce general anesthesia. These changes may be particularly pronounced in patients with heart or blood vessel disease or high blood pressure. An upper-arm cuff measures blood pressure at intervals, whereas an arterial catheter measures it continuously with each heartbeat. Although cuff measurements are widely used, their agreement with arterial catheter measurements during this period of rapid change is not fully evaluated and validated.
This observational study will include 120 patients undergoing surgery under general anesthesia at University Hospital Graz, Austria. Participants will have conditions that increase their risk of blood pressure changes and will already require an arterial catheter before anesthesia for medical reasons.
During the first five minutes after administration of the medication used to induce anesthesia, researchers will record blood pressure using both methods. The cuff will be placed on the arm opposite the arterial catheter and will take measurements at least once per minute, or more frequently if the monitor allows. Anesthesia and treatment decisions will remain the responsibility of the treating clinician. Participation does not require additional follow-up visits.
The main aim is to assess agreement in mean arterial pressure, which is the average pressure in the arteries over a heartbeat. Researchers will also compare systolic pressure, the pressure when the heart contracts, and diastolic pressure, the pressure when the heart relaxes. They will examine whether patient characteristics or treatments are associated with differences between the two measurement methods and will separately assess agreement during particularly large, rapid blood pressure changes.
The findings may help clinicians understand when cuff measurements differ meaningfully from arterial catheter measurements and inform future decisions about blood pressure monitoring during the start of anesthesia.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Intraarterial blood pressure monitoring prior to induction of anesthesia is necessary for clinical reasons
Surgery under general anesthesia
Provision of written informed consent
Less than a 5 mmHg difference between consecutive non-invasive oscillometric blood pressure measurements on the left and right arms
Disqualifiers
Inability to insert the intraarterial catheter and venous catheter on the same arm (mandatory to be on the contralateral side of the non-invasive oscillometric measurement)
Pregnancy
Planned awake fiberoptic intubation
Previous enrollment in the study
Trial population
Adults patients with a high-risk of hemodynamic changes during induction of anesthesia undergoing general anesthesia induction with clinically indicated intraarterial blood pressure monitoring before induction.
Trial design
Cohort
Prospective
Treatments tested in this trial
non-invasive oscillometric blood pressure monitoring
Other interventionIn addition to intraarterial blood pressure monitoring, non-invasive oscillometric blood pressure monitoring will be performed on the contralateral arm.
Treatment groups
Trial outcomes
Primary outcomes
Mean arterial pressure measurement Agreement between IBP and NIBP
Measurement agreement in mean arterial pressure between intraarterial (reference test) and intermittent non-invasive oscillometric (index test) blood pressure monitoring in high-risk patients having induction of anesthesia.
Secondary outcomes
Diastolic arterial pressure measurement agreement between IBP and NIBP
Measurement agreement of diastolic arterial pressure between intraarterial (reference test) and intermittent non-invasive oscillometric (index test) blood pressure monitoring in high-risk patients having induction of anesthesia.
Systolic arterial pressure measurement Agreement between IBP and NIBP
Measurement agreement of systolic arterial pressure between intraarterial (reference test) and intermittent non-invasive oscillometric (index test) blood pressure monitoring in high-risk patients having induction of anesthesia. Time Frame: In the first five minutes after induction of anesthesia
Characteristics associated with disagreement between IBP and NIBP
Patient and/or treatment characteristics associated with pairwise disagreement (\>10 mmHg) between intraarterial (clinical reference) and intermittent non-invasive oscillometric (index test) blood pressure monitoring in high-risk patients having induction of anesthesia.
Other outcomes
Subgroup Analysis in patients with severe blood pressure changes
In a predefined subgroup analysis, the primary aim and secondary aims 1 and 2 will be evaluated in situations with severe blood pressure changes (a change ≥ 50 mmHg within one minute).