About this trial
Background:
The choice whether or not to preserve the posterior cruciate ligament (PCL) in total knee arthroplasty (TKA) is coupled to the us of a PCL-retaining (CR) or a condylar (CS) insert. The CS insert is anterior-lipped (AL) to prevent anterior translation of the femur on the tibia with flexion and compensate the function of the PCL. Currently both the CR and CS insert are made of highly cross-linked polyethylene (HXPLE) to theoretically reduce wear related osteolysis. However, this also might diminish the fracture toughness and crack propagation of the insert. The investigators expect that due to the high contact forces on the anterior lip of the CS insert during flexion, especially in younger and more active patients, and the lower fracture toughness of HXPLE, the CS insert insert might show more migration, wear or other damage compared to the CR insert in the long-term, which might lead to more revisions in the CS insert compared to the CR insert.
To measure the migration and wear, during surgery tantalum markers will be inserted in the host bone using a marker inserter. The displacement of the prosthesis with reference to the host bone will be measured using model-based RSA. Both tantalum markers ande the inserter are already used for study purposes. However, the safety and usability are not registered before.
Objective:
The primary objective is to compare the migration of both the femoral and tibial component by the use of a CS insert or CR insert both made of HXPLE using model-based roentgen stereophotogrammetric analysis (mRSA). Furthermore, the safety and usability of the tantalum markers and the marker inserter will be determined. The secondary objective is to determine the influence of the type of insert on the wear, inducible displacement, survival and clinical outcomes.
Study design:
A randomized controlled trial
Study population:
Forty-four patients scheduled to undergo primary total knee replacement, aged below 70 years with an ASA-score of 1 or 2 will be needed in total, divided in two groups of 22 patients each.
Intervention:
One group receives an uncemented TKP with a CS insert, while the other group receives an uncemented TKP with a CR insert. Both will be placed using the MAKO-robotic arm using a kinematic balancing technique.
Outcomes:
Main study parameters are migration of the femoral and tibial components measured with model-based RSA software till 10 years postoperatively. Furthermore, the stability of the markers will be determined and the complications due to the markers and/or the marker inserter will be registered. The secondary parameters are wear, inducible displacement, survival, clinical outcomes and complications up to 10 years postoperatively.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Painful, disabling joint disease of the knee resulting from: non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis.
One or more compartments are involved;
Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability (meaning a varus, valgus or flexion deformity < 15 degrees);
Age between 18 and 70 years;
Disqualifiers
Any active or suspected latent infection in or about the knee joint;
Distant foci of infection which may cause hematogenous spread to the implant site;
Any mental or neuromuscular disorder which would create an unacceptable risk of prosthesis instability, prosthesis fixation failure, or complications in postoperative care;
Bone stock compromised by disease, infection or prior implantation which cannot provide adequate support and/or fixation to the prosthesis;
Trial design
Parallel
Treatments tested in this trial
Condylar Stabilizing (CS) insert
DeviceDuring TKA surgery a Condylar Stabilizing (CS) insert is implanted and tantalum beads are placed around the prosthesis.
Cruciate Retaining (CR) insert
DeviceDuring TKA surgery a Cruciate Retaining (CR) insert is implanted and tantalum beads are placed around the prosthesis.
Treatment groups
Trial outcomes
Primary outcomes
Translations (mm)
The translations of the femoral and tibial components measured in 3 axis (z, y and z) with model-based RSA over 10 years.
Rotations (degrees)
The rotations of the femoral and tibial components measured in 3 axis (z, y and z) with model-based RSA over 10 years.
Stability of the tantalum markers
Stability of the tantalum markers will be determined by the mean error value of the markers.
Safety of the tantalum markers and the inserter
Safety of the tantalum markers are determined by registration of the complications due to the markers and/or marker inserter. Complications will be expressed by numbers and percentages.
Secondary outcomes
Wear
Wear of the insert measures in 3 axis (x, y and z)
Inducible displacement (mm)
The inducible displacement gives insight into the nature and durability of the fixation and is defined as the micromotion occurring in response to external forces. The inducible displacement will be expressed by translations (x-, y- and z-direction) in mm
Inducible displacement (degrees)
The inducible displacement gives insight into the nature and durability of the fixation and is defined as the micromotion occurring in response to external forces. The inducible displacement will be expressed by rotations (x-, y- and z-direction) in degrees.
Survival
Survival of the CS and CR inserts is measured by registration of all complications and is expressed by numbers and percentages.
Sponsors and contacts