Multisensor Wireless Pressure Microcatheter For Microvascular Function Assessment In ANOCA/INOCA Patients: A Prospective, Multicenter, Single-Group Target Value Study

Trial statusRecruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
Age18+
SponsorPeking University Third Hospital

About this trial

This study aims to evaluate the sensitivity and specificity of a multi-sensor wireless pressure microcatheter for the diagnosis of CMD in patients with ANOCA/INOCA, using quantitative myocardial perfusion imaging of cardiac magnetic resonance (CMR) as a reference.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Age ≥ 18 years, gender not limited;

Symptoms of angina pectoris, or objective evidence of myocardial ischemia (e.g., electrocardiogram, myocardial injury markers, etc.);

Visual assessment of coronary angiography shows diameter stenosis < 50% and FFR > 0.8 (or cRR > 0.89);

Microvascular function assessment is planned;

Disqualifiers

Acute myocardial infarction, PCI, CABG, or valvular surgery following angiography within the past 30 days;

Severe valvular disease requiring surgical or interventional treatment;

Chest pain with known non-ischemic causes (e.g., pericarditis, pulmonary hypertension, esophageal spasm);

Contraindications to CMR (e.g., certain types of pacemakers or defibrillators, severe claustrophobia);

Trial population

This study aims to enroll patients with angina symptoms or objective evidence of myocardial ischemia, whose coronary angiography visual assessment shows a diameter stenosis of \<50% and an FFR \>0.8 (or cRR \>0.89).

Trial design

Design model

Cohort

Time perspective

Prospective

Treatments tested in this trial

Not listed

Trial groups

No trial groups listed

Trial outcomes

Primary outcomes

1

Sensitivity and specificity of multi-sensor wireless pressure microcatheter for calculating Coronary Flow Reserve (CFR) in diagnosing Coronary Microvascular Dysfunction (CMD)

Using study participants as the unit of study, and with Myocardial Perfusion Reserve (MPR) calculated by Cardiac Magnetic Resonance (CMR) quantitative myocardial perfusion imaging as the reference (MPR\<2.2 is positive, MPR≥2.2 is negative), the sensitivity and specificity of CFR calculated by multi-sensor wireless pressure microcatheter for diagnosing CMD (CFR\<2.5 is positive, CFR≥2.5 is negative) were evaluated.

Time frame
Intraoperative period (day 0) and post-operative (days 1-7)

Secondary outcomes

1

Multi-sensor wireless pressure microcatheter calculation of CFR for diagnosing CMD: accuracy, positive predictive value, negative predictive value, Receiver Operating Characteristic Curve (ROC), and Area Under the Curve(AUC)

Using study participants as the unit of study, and with MPR calculated by CMR quantitative myocardial perfusion imaging as a reference (MPR\<2.2 as positive, MPR≥2.2 as negative), the accuracy, positive predictive value, negative predictive value, ROC, and AUC of CFR calculated by multi-sensor wireless pressure microcatheter for diagnosing CMD (CFR\<2.5 as positive, CFR≥2.5 as negative) were evaluated.

Time frame
Intraoperative period (day 0) and post-operative (days 1-7)
2

Pearson correlation analysis and Bland-Altman bias analysis of CFR calculated by multi-sensor wireless pressure microcatheter and MPR calculated by CMR quantitative myocardial perfusion imaging.

Using blood vessels as the research unit, Pearson correlation analysis and Bland-Altman bias analysis were performed on the CFR calculated by multi-sensor wireless pressure microcatheter and the MPR calculated by CMR quantitative myocardial perfusion imaging.

Time frame
Intraoperative period (day 0) and post-operative (days 1-7)
3

Pearson correlation analysis of multi-sensor wireless pressure microcatheter measurements of Tmn,rest and Tmn,hyper with CMR quantitative myocardial perfusion imaging measurements of MBFrest and MBFhyper.

Using blood vessels as the unit of study, Pearson correlation analysis was performed on the Mean Transit Time in the resting state (Tmn,rest) and the Mean Transit Time in the hyperemia state (Tmn,hyper) measured by multi-sensor wireless pressure microcatheter,Myocardial Blood Flow in the resting state (MBFrest) and Myocardial Blood Flow in the hyperemia state (MBFhyper) by CMR quantitative myocardial perfusion imaging,respectively.

Time frame
Intraoperative period (day 0) and post-operative (days 1-7)
4

Multi-sensor wireless pressure microcatheter calculation of Index of Microcirculatory Resistance (IMR) for diagnosing CMD: accuracy, sensitivity, specificity, positive predictive value, negative predictive value, ROC, and AUC.

Using study participants as the unit of study, and with MPR calculated by CMR quantitative myocardial perfusion imaging as a reference (MPR\<2.2 as positive, MPR≥2.2 as negative), the accuracy, sensitivity, specificity, positive predictive value, negative predictive value, ROC, and AUC of IMR calculated by multi-sensor wireless pressure microcatheter for the diagnosis of CMD (IMR\<25 as negative, IMR≥25 as positive) were evaluated.

Time frame
Intraoperative period (day 0) and post-operative (days 1-7)