About this trial
The goal of this study is to evaluate the diagnostic performance and feasibility of non-invasive spectrophotometric haemoglobin (SpHb) measurement for detecting anaemia in infants admitted to the Neonatal Intensive Care Unit (NICU). The main questions it aims to answer are:
1. What is the agreement between non-invasive SpHb and laboratory reference total haemoglobin (tHb)? 2. How accurately can non-invasive SpHb detect clinically significant laboratory anaemia (Hb \< 10g/dL)? 3. What is the feasibility (success rate and time-to-reading) and what physiological factors (e.g., perfusion index) moderate measurement accuracy in NICU infants? Researchers will compare non-invasive continuous SpHb readings against routine invasive laboratory haemoglobin values (tHb) obtained during standard clinical care from the same clinical encounter.
Participants will:
1. Have a non-invasive optical sensor applied to an approved peripheral site (e.g., foot or palm) to record an SpHb reading whenever routine blood sampling is required by their attending clinician. 2. Undergo standard routine blood sampling (venous, arterial, or capillary) for laboratory haemoglobin testing as part of usual clinical care, with no extra blood drawn for research purposes.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Infants (preterm or term) admitted to the Neonatal Intensive Care Unit.
Gestational age ≥ 24 weeks at birth.
Admitted to NICU during the study period.
Clinical indication for laboratory haemoglobin measurement as part of routine care.
Disqualifiers
Known or suspected haemoglobinopathy.
Cyanotic congenital heart disease.
Significant skin integrity issues at the proposed sensor site (e.g. burns, severe dermatitis, wounds, oedema, or medical devices preventing probe placement).
Conditions likely to cause unreliable peripheral perfusion at the discretion of the attending clinician.
Trial design
Single group
Treatments tested in this trial
Spectrophotometry hemoglobin
Device1. The index test utilizes multi-wavelength spectrophotometry (Masimo Radical-7® with rainbow® sensors) applied topically to peripheral tissue (e.g., foot, palm, or digit). It requires zero skin punctures, creates no additional iatrogenic blood loss, and causes no pain or distress to fragile neonates. 2. Unlike standard diagnostic studies that require dedicated blood sampling for research, this protocol collects reference laboratory total haemoglobin (tHb) strictly when clinically indicated for routine care. No extra phlebotomy is performed solely for the study. 3. Non-invasive SpHb measurements are acquired within a strict 5-minute window of routine clinical blood sampling to eliminate physiological variability over time and ensure paired accuracy assessment.
Treatment groups
Trial outcomes
Primary outcomes
Agreement between non-invasive SpHb and laboratory haemoglobin measurements
Agreement between SpHb and laboratory haemoglobin will be assessed using Bland-Altman analysis, including mean bias and 95% limits of agreement. Unit of measure: g/dL
Secondary outcomes
Sensitivity and specificity of SpHb for detection of anaemia
Sensitivity and specificity of SpHb for detecting anaemia, using laboratory haemoglobin as the reference standard. Anaemia will be defined according to the pre-specified haemoglobin threshold appropriate for the infant's gestational/postnatal age. Unit of measure: Percentage of participants
Agreement between changes in SpHb and changes in laboratory haemoglobin over time
Description: The ability of serial SpHb measurements to detect changes in haemoglobin concentration over time will be assessed by comparing changes in SpHb with corresponding changes in laboratory haemoglobin measurements. Unit of measure: g/dL
Regression coefficient for perfusion index and SpHb-tHb measurement difference
Regression coefficient (β) and 95% confidence interval for perfusion index (measured by pulse co-oximeter perfusion index display, %) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant. Unit of Measure: Regression coefficient (β) per 1% change in perfusion index
Regression coefficient for gestational age and SpHb-tHb measurement difference
Regression coefficient (β) and 95% confidence interval for gestational age at birth (recorded in completed weeks from obstetric dating, per medical records) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant. Unit of Measure: Regression coefficient (β) per completed week of gestational age