About this trial
The Neuralert monitoring system, a non-invasive device worn on each wrist, is designed to provide continuous monitoring and expedited recognition of stroke by identifying asymmetry in the extremities, facilitating patient treatment and improving outcomes. This study is designed to evaluate the safety and effectiveness of the Neuralert Monitor. This study aims to show that the device produces fewer than 2 false alarms per patient per day on average, and that the device detects stroke with a sensitivity significantly greater than a device producing 2 random alerts per day.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Male or female age ≥ 22 years
Admitted or planned admission to one of the participating hospitals and followed by the Cardiothoracic (CT) Surgery or Vascular Surgery services
Intracardiac surgical or endovascular procedures, including valve replacement
Ascending aorta or aortic arch surgical or endovascular repair
Disqualifiers
Any medical or psychological conditions that, in the opinion of the investigator, would compromise the subject's safety or successful participation in the study
Baseline preoperative asymmetric upper extremity weakness as assessed by the Investigator, defined as a differential in the NIHSS upper extremity motor score between the two arms of >0
Above the wrist amputation
Unwilling to provide informed consent or no legally authorized representative willing to provide consent if the patient is unable
Trial design
Single group
Treatments tested in this trial
Neuralert Monitor
DeviceThe Neuralert monitoring system, a non-invasive device worn on each wrist, is designed to provide continuous monitoring for asymmetry in the extremities, sending an alert to healthcare workers upon identification, thus expediting recognition of stroke.
Treatment groups
Trial outcomes
Primary outcomes
Efficacy of Device
To show that the sensitivity of the Neuralert Monitor exceeds the specified performance goal, and
Safety of Device
To show that the mean number of false alarms/patient/day for the Neuralert Monitor is less than the specified performance goal.
Secondary outcomes
Length of time to detection of any stroke
Time from last known normal to detection of any stroke with weakness
Length of time to detection of stroke by Neuralert
Time from last known normal to detection by the Neuralert Monitor of any stroke with weakness
Number of alerts caused by stroke (PPV)
How often an alert is caused by confirmed stroke
Other outcomes
Number and type of causes of false alarms
A false alarm is defined as any alarm that does not result in a clinical diagnosis of stroke after the next evaluation by the clinical team
Sensitivity of device
Sensitivity of device for detection of any stroke, regardless of whether the stroke includes asymmetric weakness
Number of care visits
Number of usual care visits documented in the electronic health record
Number and type of adverse events
Any adverse events collected during the study
Sponsors and contacts