About this trial
The purpose of this study is to find out whether hearing tests performed with the Oscilla A60 audiometer, where the patient largely tests themselves with minimal staff assistance, are accurate enough to be used for diagnosing hearing loss and as the sole basis for fitting hearing aids. Adults referred for their first hearing aids at Aalborg University Hospital will have their hearing tested both with the Oscilla audiometer and by a trained audiology assistant at the same visit. Participants are then randomly assigned to receive hearing aids fitted based on either the Oscilla hearing test or the audiology assistant's hearing test. The main questions are: (1) Do the Oscilla results agree with the audiology assistant's results for both air-conduction and bone-conduction hearing thresholds? (2) Do participants fitted based on the Oscilla test report a hearing benefit that is not worse than that of participants fitted based on the standard test, two months after fitting?
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Adults (≥18 years) referred for first-time hearing aid fitting at the Department of Audiology, Aalborg University Hospital
Uncomplicated sensorineural hearing loss as defined by the Danish national clinical guidelines
Hearing thresholds exceeding 20 dB HL at two or more frequencies in the range 0.5-4 kHz
Pure-tone average (PTA-4; 0.5, 1, 2, 4 kHz) ≥30 dB HL in the better ear
Disqualifiers
Air-conduction thresholds exceeding 80 dB HL at two or more frequencies
Air-bone gap >15 dB at one or more frequencies below 1 kHz
Fluctuating hearing loss (e.g., Ménière's disease) or ongoing ototoxic medication
Ear surgery (ear canal, middle ear or inner ear) within the past 12 months
Trial design
Parallel
Treatments tested in this trial
Paired audiometry (Oscilla A60 and manual audiometry)
Diagnostic testAt Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (\<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order.
Oscilla-based hearing aid fitting
DeviceBilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the Oscilla A60 audiogram as the sole audiometric basis, verified with auto-REM (within ±5 dB of target). Other names: Oscilla A60 Diagnostic Audiometer.
Manual audiometry-based hearing aid fitting
DeviceBilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the audiology assistant audiogram, verified with auto-REM (within ±5 dB of target). Standard clinical practice.
Treatment groups
Trial outcomes
Primary outcomes
Agreement between Oscilla A60 and manual audiometry for air-conduction thresholds
Bland-Altman mean difference (Oscilla minus manual) and 95% limits of agreement in dB, plus mean absolute difference, at 0.5, 1, 2 and 4 kHz separately. Agreement is acceptable if the 95% CI of the bias is within ±5 dB and the 95% limits of agreement are within ±15 dB. Both ears included, clustering handled with a mixed-effects model.
Agreement between Oscilla A60 and manual audiometry for bone-conduction thresholds
As above for bone-conduction thresholds at 0.5, 1, 2 and 4 kHz. Agreement is acceptable if the 95% CI of the bias is within ±5 dB and the 95% limits of agreement are within ±20 dB.
Change in SSQ12 total score
Speech, Spatial and Qualities of Hearing Scale, 12-item version; total score = mean of 12 items, range 0-10, higher = better. Non-inferiority if the lower bound of the two-sided 95% CI for the between-group difference in change (Oscilla minus control) is above -0.81 points.
Secondary outcomes
IOI-HA total score
International Outcome Inventory for Hearing Aids, Danish version; 7 items scored 1-5, total range 7-35, higher = better outcome. Compared between groups
Daily hearing aid use
Mean daily hearing aid usage time in hours/day from the hearing aid data log, compared between groups
Change in SSQ12 subscale scores
Change in the speech, spatial and qualities subscale scores (each range 0-10, higher = better)