About this trial
Background: Recommendations for peri-operative management in patients with implanted cardiac electronic devices (CIED) are often based on older data from case reports and small collectives.
Objective: To evaluate the peri-operative management and outcome of patients with implanted CIED undergoing non-CIED related surgery or catheter-interventional procedures in clinical routine.
Study design: bi-center, non-randomized, observational registry, retrospective data-collection, on-going prospective patient enrollment, descriptive statistics
Primary endpoint: number and type of peri-operative adverse device related events (ADE) Secondary endpoints: pre-interventional data (patient characteristics, data from CIED interrogation); peri-interventional data (type of surgery/intervention, anesthesiology techniques, any AE), postinterventional data (data from post-interventional CIED interrogation, need for reprogramming / device revision).
Inclusion criteria: implanted CIED, non-CIED related surgical or catheter-based intervention, peri-procedural CIED interrogation, age \>18 years Exclusion criteria: no implanted CIED, no data from any peri-procedural CIED interrogation available
Patient enrollment: retrospectively beginning from 2008, further on-going prospective inclusion Sample size: For the observational study, there is no pre-specified sample size. Data from more than 500 patients undergoing \> 700 interventions are expected.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
implanted CIED (e.g. pacemaker, ICD, CRT),
performed (non-CIED related) surgical or catheter-based procedure,
data from peri-procedural CIED interrogation available
age >18 years
Disqualifiers
no implanted CIED,
no data from any peri-interventional CIED interrogation available
Trial population
Enrollment of all consecutive patients according to the inclusion criteria, undergoing surgery or catheter-ablation at the Klinikum Fuerth or Klinikum Nueremberg. Data collection will be * retrospectively beginning from 2008, * on-going with further prospective data collection and patient enrollment. If possible, patients will be enrolled at pre-interventional device interrogation. If there is no pre-interventional device interrogation (e.g. in case of urgent or emergency surgery), patients will be enrolled retrospectively.
Trial design
Cohort
Prospective
Treatments tested in this trial
surgery or catheter based cardiac procedure (catheter ablation)
Procedure/Surgerypatients with implanted CIED (pacemaker, ICD or CRT) that undergo any surgery or any catheter based cardiac procedure (e.g. radiofrequency ablation)
Treatment groups
Trial outcomes
Primary outcomes
surgery or catheter-based procedure related (S)ADE
any (S)ADE related to the surgical or catheter-based procedure
Secondary outcomes
device: post-procedural pacing threshold increase
post-procedural pacing threshold increase \> 50% compared with the pre-procedural threshold (given a safety margin of 100%)
device: post-procedural sensing decrease
post-procedural sensing decrease by \> 50% compared with the pre-procedural sensing
device: post-procedural lead impedance
post-procedural lead impedance is different by \> 25% compared with the pre-procedural impedance
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Klinikum-Fuerth
Lead sponsor
Klinikum Nürnberg
Collaborator