Atrial Shunt for the Treatment of Heart Failure With Reduced Ejection Fraction

Trial statusRecruiting
Trial phasePhase 2, Phase 3
Trial typeInterventional
Biological sexAll
Age18+
SponsorBeijing Anzhen Hospital

About this trial

To evaluate the safety and effectiveness of the Atrial Shunt for the treatment of heart failure with reduced ejection fraction (HFrEF)

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

1. Age ≥ 18 years 2. Diagnosed with HFrEF and receiving Guideline-Directed Medical Therapy (GDMT) for at least 1 month 3. Echocardiographic LVEF between 20% and 40% (20 ≤ LVEF ≤ 40%) 4. Presence of HF-related symptoms and signs, NYHA Class III 5. At least one HF-related hospitalization in the past year or NT-proBNP ≥ 600 pg/mL (sinus rhythm) / ≥ 1000 pg/mL (atrial fibrillation) 6. Invasive hemodynamics: mLAP or mPCWP ≥ 15 mmHg, or left-to-right mean pressure gradient ≥ 5 mmHg 7. Fully understand the study design and follow-up schedule, agrees to participate, and voluntarily signs the Informed Consent Form

Disqualifiers

1. End-stage heart failure on the heart transplant waiting list (including meeting LVAD indications) 2. Three consecutive non-invasive systolic blood pressure readings < 90 mmHg 3. Primary structural valvular disease currently indicating surgery or interventional therapy, or severe coronary artery disease requiring revascularization 4. ECG QRS duration ≥ 130 ms with high risk of sudden cardiac death without therapeutic intervention (including meeting CRT/CRT-D indications) 5. Significant right ventricular (RV) dysfunction or severe pulmonary hypertension. Defined by clinical evidence of hepatic congestion, moderate-to-severe edema, or abdominal distension accompanied by poor appetite; significant right atrial or RV enlargement/dysfunction; moderate-to-severe tricuspid regurgitation; mean right atrial pressure > 10 mmHg; systolic pulmonary artery pressure > 60 mmHg; or pulmonary vascular resistance > 3 Wood Units 6. Myocardial infarction or cardiac intervention within the past 1 month 7. Stroke within the past 1 month (excluding lacunar infarction) 8. Decompensated liver cirrhosis (Child-Pugh Class B or worse), or severe chronic pulmonary disease requiring oxygen therapy 9. Anatomical abnormalities precluding procedure completion or making the patient unsuitable for procedure 10. Atrial thrombosis 11. Contraindication to antithrombotic therapy 12. Life expectancy < 1 year 13. Uncontrolled active infection 14. Pregnant, lactating, or women of childbearing potential planning pregnancy in the near future 15. Currently participating in another drug or medical device clinical trial 16. Investigator deems the patient unsuitable for this trial

Trial design

Design model

Parallel

Treatments tested in this trial

  • D-shant Atrial Shunt

    Device

    The D-Shant atrial shunt device is a novel device therapy for heart failure, classified as a non-active implantable cardiovascular device. It is manufactured using a braided Nitinol technique, thermally set via assembly molds to form a waist with a fixed-orifice support and a hollow double-disc structure. Braided wires reinforce waist support and maintain a circular orifice shape. Once implanted, it forms a stable interatrial shunt. The device is delivered via the femoral vein to the interatrial septum, where the shunt orifice facilitates blood flow diversion and pressure decompression. Various models with different waist and disc diameters are available to accommodate varying degrees and etiologies of heart failure.

  • Guideline-Directed Medical Therapy (GDMT) for heart failure with reduced ejection fraction

    Drug

    Guideline-Directed Medical Therapy (GDMT) for heart failure with reduced ejection fraction

Treatment groups

208 Participants
are divided into 2 treatment groups
Group A: Device group (Interventional group)Experimental treatment 2 interventions
Group B: no Device group (Control Group)Active comparator 1 intervention

Trial outcomes

Primary outcomes

1

A composite event of all-cause mortality, heart transplantation or left ventricular assist device implantation due to heart failure, heart failure-related rehospitalization post-procedure and Unplanned outpatient worsening heart failure within 12 months

A composite event of all-cause mortality, heart transplantation or left ventricular assist device implantation due to heart failure, heart failure-related rehospitalization post-procedure and Unplanned outpatient worsening heart failure within 12 months. Hospitalization for heart failure requires an admission to an in-patient unit or an emergency room stay for ≥24 hours (or \<24 hours if subject dies in the emergency room). - Outpatient worsening heart failure requires an unplanned visit to a doctor's office, urgent care center or emergency room visit with stay \<24 hours. - Worsening heart failure must be present for both conditions, the definition of which requires all three of the following to be present: 1. Deterioration of HF symptoms: at least 1 of the following symptoms 2. Deterioration of HF signs: 2 physical examination findings or 1 physical examination + 1 laboratory or invasive finding 3. Urgent escalation of therapy

Time frame
From enrollment to the end of treatment at 12 months

Secondary outcomes

1

6-minute walk distance (6MWD) (0-1000m, higher scores indicate a better outcome)

Change in exercise capacity post-procedure: 6-Minute Walk Test

Time frame
From enrollment to the end of treatment at 12 months
2

New York Heart Association (NYHA) Functional Classification (I-IV, higher scores indicate a worse outcome)

Change in New York Heart Association (NYHA) Functional Classification

Time frame
From enrollment to the end of treatment at 12 months
3

Change in NT-proBNP levels

Change in NT-proBNP levels

Time frame
From enrollment to the end of treatment at 12 months
4

Left atrial diameter

Echocardiographic Left atrial diameter

Time frame
From enrollment to the end of treatment at 12 months

Other outcomes

1

Device-Related Adverse Events

Cardiac tamponade Air embolism Systemic embolization syndrome Thrombus formation on the device surface Device malfunction requiring emergency surgical removal

Time frame
From enrollment to the end of treatment at 12 months

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Beijing Anzhen Hospital

Lead sponsor

Union Hospital, Tongji Medical College, Huazhong University of Science and Technology

Collaborator