About this trial
Coronary angiography (CAG) for diagnostic or therapeutic purposes such as percutaneous coronary intervention (PCI) is one of the common procedures which require the use of intravenous contrast media. The reported incidence of contrast induced nephropathy (CIN) in high-risk patients following CAG varies from 10% to 30%. The high rate of CIN in post-PCI patients could be related either to the patient (advanced age, previous CKD, diabetes, dehydration, and concomitant use of other nephrotoxic drugs) or procedure related (intra-arterial route of administration, use of high osmolar contrast media, repeated exposure to contrast within 48 hours, volume of contrast used). Several strategies to prevent or treat CIN have been developed, including hydration, N-acetyl-cysteine, statins, ascorbic acid, bicarbonate, aminophylline, forced diuresis, renal replacement therapy, and choice of low-osmolarity or alternative agents, but one of the most obvious means is to minimize contrast volume. The DyeVert plus Contrast Reduction System, is designed to lower the amount of contrast dye the kidneys are exposed to during a procedure. Because the amount of contrast dye is precisely controlled. The purpose of this prospective study is to understand how the monitoring system of Dye-Vert Plus will impact Acute Kidney Injuries rates in high-risk patients undergoing cardiac catheterization when used in conjunction with a standardized hydration policy.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
18 years of age or older
Scheduled to undergo CAG and/or PCI
Baseline estimated glomerular filtration rate (eGFR) of ≥20 and ≤60 mL/ min/1.73 m2
Serum creatinine > 1.5mg/dl
Disqualifiers
91 years of age or older
Serum creatinine < 1.5mg/dl
eGFR > 60ml/min
Pregnancy
Trial design
Parallel
Treatments tested in this trial
DyeVert Plus System
DeviceThe DyeVert Plus System also contains a reusable Contrast Monitoring Wireless Display (CMW) which communicates with the Dye-Vert Plus Disposable to allow real-time monitoring and display of contrast volumes manually injected compared to a predefined, physician-entered contrast usage threshold throughout the procedure. At the end of the procedure, the CMW displays total procedure contrast volume used (mL), % of physician specified limit , total procedure contrast volume saved (mL), and % contrast saved
Standardized hydration protocol
Other interventionhydration protocol based up on the Poseidon protocol
Treatment groups
Trial outcomes
Primary outcomes
Monitoring of AKI
Determined by GFR levels
Sponsors and contacts