About this trial
The objective of this study is to verify the form, fit and function of the Philips FAST technology with Masimo neonatal, infant and pediatric sensors for SpO2 monitoring in the respective indicated populations. The study design uses convenience sample data collection from neonatal, infant, and pediatric sensors within their respective indicated patient populations that have a range of different skin pigmentations, simultaneously with an FDA cleared pulse oximeter as a comparator.
Eligibility criteria
This trial accepts healthy volunteersQualifiers
Subject falls within the indicated population for the Philips FAST SpO2 Monitor with the Masimo sensor.
Subject is less than 22 years of age (pediatric subject per US FDA definition).
Disqualifiers
Subject has a skin condition or deformity at the planned sensor application site, which would preclude sensor placement and measurements as standard of care (e.g., psoriasis, eczema, angioma, scar tissue, burn, fungal infection, substantial skin breakdown).
Confounders of pulse oximetry per sensor direction for use (DFU).
Subject with distinct geographic variances in skin pigmentation (e.g., vitiligo), where the sensor is applied.
Subject has an absence or deformities of limbs or severe edema, which would interfere with sensor application or prevent the proper fit of the sensors.
Trial design
Single group
Treatments tested in this trial
Masimo Sensor (RD SET and/or LNCS) connected to Philips Monitor using Philips FAST SpO2 Algorithm
DeviceSubjects may use multiple Masimo sensors (RD SET and/or LNCS). All subjects will have at least one Masimo sensor (RD SET or LNCS) connected to Philips Monitors using Philips FAST SpO2 Algorithm
Treatment groups
Trial outcomes
Primary outcomes
The acceptability of the quality of SpO2 signal obtained during non-invasive oxygen saturation monitoring using Philips FAST SpO2 technology with Masimo neonatal, infant, and pediatric sensors. Accuracy root mean square (ARMS) will be calculated.
Secondary outcomes
The acceptability of the fit of sensor by visual observation of the emitter and detector alignment.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.