Pivotal Study to Assess Safety and Performance of Neola®, a Novel Lung Monitoring Device for Neonates

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age1-44
SponsorNeola Medical Inc

About this trial

Pivotal study to assess safety and performance of Neola®, a novel lung monitoring device for neonates

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Preterm or term born neonates with gestational age between 26 and 40 weeks

Age between 1 day post-natal and 44 weeks corrected gestational age

Weight between 1000 g and 3500 g

Patient treated at a neonatal intensive care unit.

Disqualifiers

Known cardiopulmonary congenital anomalies

Patients with trisomies or other chromosomal abnormality

Patients not expected to survive

Non-intact skin or other skin conditions (such as skin lesions) that do not allow the use of skin adhesives. (e.g., Bullous impetigo, Staphylococcal scalded skin syndrome, localized lesions of herpes simplex virus)

Trial design

Design model

Single group

Treatments tested in this trial

  • Lung monitoring with the Neola device

    Device

    Lung monitoring with the Neola device

Treatment groups

22 Participants
are divided into 1 treatment group
Group A: Arm 1Experimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

The number of device related events including skin reactions

The primary objective of the study is to evaluate safety with regards to adverse events (AE) and in particular device related adverse events. Outcome measure: The number of device related events including skin reactions.

Time frame
2 days
2

The number of serious and non-serious adverse events occurring during the trial

The primary objective of the study is to evaluate safety with regards to adverse events (AE) and in particular device related adverse events. Outcome measure: The number of serious and non-serious adverse events occurring during the trial.

Time frame
2 days

Secondary outcomes

1

The device's ability to continuously measure and present the clinical parameters while the neonates are in a resting state/sleeping

This endpoint will be presented as the percentage of the total resting/sleeping time that the device is able to measure the clinical parameters relative lung volume and absolute oxygen concentration.

Time frame
7 days
2

The device's ability to continuously measure and present the clinical parameters during normal handling of the neonates

This endpoint will be presented as the percentage of the total handling time that the device is able to measure the clinical parameters relative lung volume and absolute oxygen concentration.

Time frame
7 days
3

The adhesive performance of the probes in terms of ability to be repositioned and stay attached

The adhesive performance of the probes in terms of ability to be repositioned and stay attached. This exploratory endpoint will be evaluated the percentage of probe adhesives that were able to stay attached for the duration of the measurement session

Time frame
7 days

Other outcomes

1

Evaluate the correlation between Neola® measured clinical data and clinical data available from other monitoring devices, procedures or tests.

The endpoint will be valuated using diagnostic methods such as, but not limited to: peripheral oxygen saturation (SpO2), x-ray images, Blood gas analysis data (partial pressure of oxygen (PaO2) and carbon dioxide (PaCO2)), respiratory support settings, respiratory rate, heart rate, blood pressure.

Time frame
7 days

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

This trial is not recruiting at the moment. You can still explore other options: