About this trial
This study will test the control and measurements of a new system for airway support in healthy volunteers. The goals of the study are to evaluate how accurately the system can stabilize airway pressure to a desired level, evaluate the accuracy of the system's reported volume of each breath (tidal volume), evaluate the accuracy of the system's reported respiratory rate, evaluate the accuracy of exhaled carbon dioxide monitoring with a standard carbon dioxide monitor when the novel PAP system is used, and evaluate general comfort of the volunteers while using the system.
Eligibility criteria
This trial accepts healthy volunteersQualifiers
Healthy volunteers
Disqualifiers
Have any respiratory or cardiovascular conditions
i.e., cold, allergies, congestion, pneumonia, lung cancer, COPD, acute respiratory distress syndrome, etc.
Trial design
Single group
Treatments tested in this trial
CPAP
DeviceAirway pressure support through the investigational device.
Treatment groups
Trial outcomes
Primary outcomes
Tidal Volume Accuracy
The mean and standard deviation of the percent difference in tidal volume measurement between our PAP system and the respiratory monitor
Secondary outcomes
Respiratory Rate
The average respiratory rate reported by the respiratory monitor and rate reported by the PAP system for each volunteer will be compared. The average difference for each volunteer will be calculated, and a 95% confidence interval will be constructed, with the ideal mean difference being zero with low variance.
EtCO2
End-tidal carbon dioxide measurements at 10 cmH2O of mask pressure will be compared between when the system turned off flow during exhalation and when delivering flow throughout the entire breath. We anticipate that end-tidal CO2 levels will read lower when flow is delivered continuously during exhalation because of dilution of the expired gas, and this will be demonstrated by calculating the average difference in end-tidal CO2 between these two approaches.
Mask Pressure Control Accuracy
The measured pressure in the mask compared to the set pressure in the 50 milliseconds immediately preceding inhalation will be assessed for evaluating pressure control. Average differences at each set pressure will be calculated. We anticipate these numbers to be less than 1 cmH2O.
Locations
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
University of Utah
Lead sponsor
Dynasthetics LLC
Collaborator