About this trial
The purpose of this clinical trial is to evaluate the safety and efficacy of the Prizvalve Pro™ transcatheter aortic valve system in the treatment of patients with severe aortic stenosis who are at high risk of surgery or who are not suitable for surgery.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Age ≥ 65 years ;
Patients with symptomatic severe aortic valve stenosis (evaluated by echocardiography: transaortic valve pressure gradient ≥40mmHg (1mmHg=0.133kPa), or transaortic valve blood flow rate ≥4m/s, or aortic Valve area<1.0cm², or AVA <0.5cm² /m²);
Cardiac function NYHA ≥ II;
Life expectancy> 12 months;
Disqualifiers
Acute myocardial infarction occurred within 1 month before this treatment;
Congenital unicuspid aortic valve;
Any therapeutic cardiac operation that is performed within 30 days of the index procedure (Previous PCI or implantation of a permanent pacemaker are not considered exclusion criteria);
Patients with other prosthetic heart valves, prosthetic rings already implanted in the heart, or severe mitral valve closure insufficiency (>3+);
Trial design
Single group
Treatments tested in this trial
Prizvalve Pro™ transcatheter aortic valve system
DeviceTranscatheter Aortic Valve Replacement
Treatment groups
Trial outcomes
Primary outcomes
Composite endpoint event rate at 30 days
Composite endpoint event includes all-cause death, stroke, myocardial infarction, permanent pacemaker implantation, serious vascular and access-related complications, surgical intervention for aortic valve reoperation, valve-in-valve surgery.
Secondary outcomes
Device success rate
Device success is defined as 1. The device enters the vascular approach, be transported, released and implanted successfully, and the delivery system is removed from the body successfully; 2. The implanted valve achieves the desired efficacy (mean transvalvular pressure difference \<20 mmHg or peak flow velocity \<3 m/s in the aortic valve, no severe prosthetic aortic regurgitation or perivalvular leak).
Procedural success rate
Procedure success is defined as 1. Implantation of ≤2 valves, with the implanted valves placed in the appropriate anatomic position; 2. No intraoperative or immediate postoperative mortality; 3. The intended purpose of the procedure was achieved without serious intraoperative complications (coronary occlusion, septal perforation, mitral valve damage or incompetence, pericardial tamponade, surgical termination, and open-heart surgical treatment for cardiac surgery).
Delivery system and component performance evaluation
Delivery system performance evaluation includes delivery, transvalve, release positioning, balloon filling and retrieval, visibility, and retrieval performance. (For each assessment: 1-4 points;1=excellent, 2=good, 3=fair, 4=poor).
Valve function
Valve function includes aortic orifice area, transvalvular pressure difference, valvular regurgitation, and perivalvular leakage.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Shanghai NewMed Medical Co., Ltd.
Lead sponsor
The General Hospital of Northern Theater Command
Collaborator
West China Hospital
Collaborator