About this trial
This observational study aims to identify the incidence and importance of discrepancies between measurements of intraoperative invasive and noninvasive blood pressure monitoring in patients undergoing non-cardiac surgeries under general anesthesia . The main questions it aims to answer are :
* The importance of the differences, in mmHg, between the non-invasive and invasive blood pressure measurements (NIBP-IBP) on systolic, diastolic and mean arterial pressure. * Identify the predictive factors associated with these differences.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Fully consented, adult patients above 18 years old;
Undergoing surgery of duration time expected at least 60 minutes using general anesthesia;
Supine positioning during the surgery;
Surgery requiring an arterial line;
Disqualifiers
Atrial fibrillation, multifocal atrial tachycardia, or any other irregular hear rythm;
Gradient of mean arterial pressure between the two arms greater than 5 mm Hg, as measured during the recruitment process.
Trial population
Adult patients undergoing non-cardiac elective surgery expected to last at least 60 minutes, under general anesthesia, and to require an arterial line.
Trial design
Cohort
Prospective
Treatments tested in this trial
Not listed
Trial groups
Trial outcomes
Primary outcomes
Difference between non-invasive blood pressure (NIBP) and invasive blood pressure (IBP) measurements
Describe the difference, in mm of Hg, between intraoperative monitoring of noninvasive and invasive blood (NIBP-IBP) on systolic, diastolic and mean arterial pressures
Predictive factors associated with the between non-invasive blood pressure (NIBP) and invasive blood pressure (IBP) measurement differences
Identify predictive factors (such as patient, surgery, and anesthesia characteristics) statistically associated with the appearance of NIBP-IBP measurement differences
Secondary outcomes
Comparison of preoperative NIBP and pre-induction NIBP
Compare preoperative non-invasive blood pressure (NIBP) measurements prior to the surgery and the average pre-induction NIBP recorded in the operating room
Compare the discrepancies between non-invasive blood pressure (NIBP) and invasive blood pressure (IBP) measurements to the accuracy standards of the Association for the Advancement of Medical Instrumentation
Compare the NIBP-IBP discrepancy to the accuracy standards of the Association for the Advancement of Medical Instrumentation (AAMI): a maximal mean difference of 5 mm Hg and a standard deviation of 8 mm Hg.
Impact of patient comorbidities (age, weight, height, BMI, heart condition, vascular condition) on occurrences of discrepancies between non-invasive blood pressure (NIBP) and invasive blood pressure (IBP) measurements
Evaluate the impact of patient's characteristics and comorbidities collected via REDCap questionnaire (age, weight, height, BMI, heart condition, vascular condition) on the occurrence of NIBP-IBP discrepancies.
Statistical correlation between radial artery ultrasound measurements on the occurrence of discrepancies between non-invasive blood pressure (NIBP) and invasive blood pressure (IBP) measurements
Evaluate the impact, via statistical analysis, of radial artery ultrasound characteristics (inter-intima diameter, velocity) on the occurrence of NIBP-IBP discrepancies.