About this trial
The present study, OBPM\_Calfree2024, with N = 85 participants minimum over 1 visit (lasting around 1h), has been designed to collect raw optical data with Aktiia.product-G2C Clinical investigational system together with several reference systems within a cohort of subjects characterized by a wide variety of phenotypes.
Eligibility criteria
This trial accepts healthy volunteersQualifiers
Subjects aged 21 to 85yo
Subjects fluent in written and spoken French
Subjects agreeing to attend the study visit and follow study procedures
Subjects that have signed the informed consent form.
Disqualifiers
Amputated index fingers
Damaged/injured skin at index fingers
Damaged/injured skin at wrists
Subjects suffering from sustained cardiac arrhythmias that can lead to weak or unstable pressure pulses including tachycardia (heart rate at rest > 120bpm) and atrial fibrillation
Trial design
Single group
Treatments tested in this trial
Investigational device Aktiia G2C
DeviceStudy subjects are asked to stay seated and relaxed while successive measurements are taken with Aktiia G2C.
Marketed device used as reference for blood pressure monitoring: double auscultation cuff
DeviceStudy subjects are asked to stay seated and relaxed while successive measurements are taken with cuff for double auscultation.
Marketed device used as comparative device for blood pressure and pulse rate monitoring: Aktiia Bracelet G2
DeviceStudy subjects are asked to stay seated and relaxed while successive measurements are taken with Aktiia Bracelet G2.
Marketed device used as comparative device for blood pressure and pulse rate monitoring: oscillometric upper-arm cuff
DeviceStudy subjects are asked to stay seated and relaxed while taking successive measurements with an oscillometric upper-arm cuff.
Marketed device used as comparative device for blood pressure and pulse rate monitoring: oscillometric wrist cuff
DeviceStudy subjects are asked to stay seated and relaxed while taking successive measurements with an oscillometric wrist cuff.
Marketed device used as reference for pulse rate monitoring: pulse finger oximeter
DeviceStudy subjects are asked to stay seated and relaxed while successive measurements are taken with pulse finger oximeter.
Treatment groups
Trial outcomes
Primary outcomes
Number of study subjects that contribute with analyzable raw data collected
Min 85 study subjects contribute with analyzable raw data collected
Secondary outcomes
Blood pressure mean value of differences
The mean value of the differences between Aktiia G2C and double auscultation blood pressure measurements.
Blood pressure standard deviation of differences
The mean value of the differences between Aktiia G2C and double auscultation blood pressure measurements.
Pulse rate root-mean-square error
The root-mean-square difference between the Aktiia G2C pulse rate determinations and the reference method.