About this trial
This is a prospective, randomized, stratified clinical trial by gestational age with 5 arms including preterm newborns (PTNBs) with a gestational age (GA) ≥ 28 weeks, in a public hospital in northeastern Brazil.
* There is no difference in efficacy and safety between the 5 interfaces. * To describe the differences in efficacy and safety of the 5 interfaces. After signing the informed consent form, stratification will be performed by gestational age (≥ 28 weeks and \< 32 weeks, and ≥ 32 weeks and \< 37 weeks). After respiratory assessment, if mild to moderate respiratory distress (BSA between 1 and 6) is confirmed. Following randomization for allocation to the groups, eligible newborns who had multiple births will be randomized individually. Those who use mechanical ventilation and who will be extubated will be randomized following the same inclusion criteria.
Patients will be randomly allocated into 5 groups according to the NIV interface:
* Group 1 (G1) - Infant Flow LP® Interface; * Group 2 (G2) - RAM® Cannula; * Group 3 (G3) - Dräger Babyflow® Nasal Mask; * Group 4 (G4) - Conventional binasal prongs (Ventcare®); * Group 5 (G5) - Fanem® Flexible Anatomical Binosal Prongs.
Randomization will be done using sequentially numbered, sealed, opaque envelopes containing the group allocation, separated into two boxes as per the stratification described above. Vital signs and blood pressure will be collected before the use of ventilatory support on a specific data collection form and will continue at 4-hour intervals at the following times (8:00, 12:00, 16:00, 20:00, 0:00, 4:00) until non-invasive ventilation is withdrawn.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Preterm infants (PTIs) with a gestational age of ≥ 28 weeks admitted to the Neonatal ICU;
PTIs with mild to moderate respiratory distress (identified by the Silverman-Andersen score);
Participation in the study authorized by parents or legal guardians via the signing of an informed consent form.
Disqualifiers
Patients with multiple congenital malformations;
Patients with intolerance to the use of NIV interfaces. Intolerance to interface use is defined as: psychomotor agitation, a ≥ 20% change in vital signs (increase or decrease) relative to the infant's baseline values, and/or a drop in peripheral oxygen saturation (SpO2) below the target range for the infant.
Trial design
Parallel
Treatments tested in this trial
Infant Flow
DeviceInfant Flow LP non-invasive ventilation interface
Ram Cannula
DeviceRam Cnnula non-invasive ventilation interface
Baby Flow nasal mask
DeviceBaby flow nasal mask non-invasive ventilation interface
Conventional binasal prongs
DeviceConventional binasal prongs non-invasive ventilation interface
Fanem Flexible Anatomical Binaural Prongs
DeviceFanem Flexible Anatomical Binaural Prongs non-invasive ventilation interface
Treatment groups
Trial outcomes
Primary outcomes
Incidence of Non-Invasive Ventilation (NIV) Failure
NIV failure is reported as a single composite binary outcome (Yes/No per participant) defined as the occurrence of at least one of the following standardized clinical criteria: Inspired oxygen fraction (FiO2) \> 60%; Positive end-expiratory pressure (PEEP) \> 8 cmH2O; Peak inspiratory pressure (PIP) \> 20 cmH2O or required pressure delta ≥ 12 cmH2O; Increase or lack of reduction in respiratory rate within 2 hours after NIV initiation (without fever or pain); Silverman-Andersen Score (SAS) \> 6, or recurrent apneas; SpO2/FiO2 ratio ≤ 180, or blood gas analysis with pH \< 7.22; Altered behavioral state assessed by the adapted Brazelton scale (states 5 or 6); Hemodynamic instability requiring clinical intervention; Severe cardiac arrhythmias or persistent bradycardia; Severe abdominal distension requiring NIV interruption;Worsening of underlying disease associated with deterioration of vital signs; Grade 3 nasal injury and/or upper lip facial injury score ≥ 7.
Secondary outcomes
Compare the effectiveness of the 5 VNI interfaces
Incidence of Non-Invasive Ventilation (NIV) Failure • Number of preterm infants who experience NIV failure, defined as the need for endotracheal intubation or escalation of ventilatory support according to protocol criteria.
Incidence and Severity of Interface-Related Skin Injuries
Incidence of skin and mucosal injuries caused by the interfaces or fixation devices. Severity is classified into stages: Stage 1 (non-blanchable erythema), Stage 2 (superficial erosion/partial-thickness skin loss), and Stage 3 (tissue necrosis).
Preterm Infant Behavior During Ventilatory Support
The newborn's level of alertness was assessed using an adapted version of the Brazelton Neonatal Behavioral Assessment Scale. This assessment classifies the newborn's behavioral state on a six-point ordinal scale, ranging from 1 (deep sleep) to 6 (agitated alertness). These represent categories within a single assessment, aimed at measuring the newborn's behavioral state.
NIV Modes, Escalation, De-escalation, and Ventilatory Parameters
Record of NIV modes applied, escalation/de-escalation events, and parameters including PEEP, PIP (cmH2O), Inspiratory time, Rise time (seconds), Flow (L/min), and FiO2 (%).
Sponsors and contacts