Safety and Effectiveness of TaurusTrio™ Heart Valve System in Patients With Severe Aortic Regurgitation (AR)

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorPeijia Medical Technology (Suzhou) Co., Ltd.

About this trial

To evaluate the safety and effectiveness of the TaurusTrio™ Heart Valve System in a patient population with symptomatic severe AR requiring replacement/repair of their native aortic valve that are at high risk for open surgical aortic valve replacement/repair (SAVR).

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Patients who voluntarily participate and sign the informed consent and are able to comply with the entire trial process;

Age ≥ 18 years;

Jet width ≥ 65% of LVOT

Vena contracta width of > 6 mm

Disqualifiers

Congenital uni- or bicuspid aortic valve morphology;

Previous prosthetic aortic valve (bioprosthesis or mechanical) implant;

Mitral regurgitation or Tricuspid regurgitation> moderate;

Clinically significant coronary artery disease (CAD) requiring revascularization within 30 days prior to index procedure, or planned CAD revascularization procedure within 12 months after index procedure;

Trial design

Design model

Single group

Treatments tested in this trial

  • TaurusTrio™ Heart Valve System

    Device

    The TaurusTrio™ Heart Valve System contains the following sub-components: a prosthetic transcatheter porcine pericardial aortic valve, 20Fr Introducer Sheath System, transfemoral Delivery Catheter, and Loading Tool.

Treatment groups

116 Participants
are divided into 1 treatment group
Group A: A single set of testExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

All-cause mortality at 12 months

All-cause mortality

Time frame
12 months

Secondary outcomes

1

Improvement in quality of life (QoL) at 12 months over baseline, as measured by the Kansas City Cardiomyopathy Questionnaire (KCCQ).

Improvement in quality of life (QoL)

Time frame
12 months
2

Composite safety endpoint at 30 days

all-cause death;all stroke; life-threatening or major bleeding;acute kidney injury (AKI) stage 2, 3 or dialysis;major vascular complications;surgery/intervention related to the device (including coronary intervention);permanent pacemaker implantation;moderate or severe total aortic regurgitation

Time frame
30 days after procedure

Other outcomes

1

Technical Success at exit from OR, hybrid room or cath lab post-index procedure

Absence of procedural mortality; and Successful access, delivery, and retrieval of transcatheter delivery system; and Deployment and correct positioning of a single intended THV; and Freedom from re-intervention related to the device or access procedure

Time frame
Immediate after procedure
2

Device Success at 1-month consisting of

Absence of procedure mortality; and Correct position of a single prosthetic heart valve Intended performance of the prosthetic heart valve (i.e., no prosthesis mismatch and mean aortic valve gradient \<20 mmHg or peak velocity \<3 m/s; and no moderate or severe prosthetic valve regurgitation) Intended performance of the prosthetic heart valve (i.e., no prosthesis mismatch and mean aortic valve gradient \<20 mmHg or peak velocity \<3 m/s; and no moderate or severe prosthetic valve regurgitation)

Time frame
30 days after procedure
3

Procedural Success at 1-month consisting of

Device success at 30 days; and No device- or procedure-related serious adverse events (defined as below): i) No disabling or non-disabling stroke ii) No life-threatening bleeding (\>4 units PRBC) iii) No acute kidney injury - Stage 2 or 3 (including dialysis) iv) No major vascular complication v) No myocardial infarction or cardiac ischemia requiring PCI or CABG vi) No valve-related dysfunction requiring repeat procedure (BAV, TAVI, or SAVR)

Time frame
30 days after procedure

Sponsors and contacts