Safety and Effectiveness of the HIT Reverse HRS

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age50-75
SponsorHip Innovation Technology

About this trial

The purpose of this study is to determine if the use of the HIT Reverse Hip Replacement System is safe and effective in patients undergoing total hip replacement. We will determine whether it is safe and effective by comparing it to the control devices.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Patient requires primary THA due to NIDJD or any of its composite diagnoses such as osteoarthritis, avascular necrosis, posttraumatic arthritis, slipped capital epiphysis, fracture of the pelvis, and diastrophic variant requiring unilateral primary THA.

Patient is between 50 (inclusive) and 75 (inclusive) years of age at the time of enrollment.

Patient has preoperative medical clearance and is free from or treated for medical conditions that would pose excessive operative risk.

Patient has a signed and dated Informed Consent Form (ICF).

Disqualifiers

Patient has received earlier THA, hemi-arthroplasty, or fusion in the contralateral hip in the last 24 months.

Patient has received earlier THA, hemi-arthroplasty, or fusion in the contralateral hip more than 24 months ago and has a contralateral hip-specific pain rating of ≥2 on a Numeric Rating Scale 0-10.

Patient needs bilateral hip replacement or has a planned THA on the contralateral hip joint in the next 24 months.

Patient needs knee arthroplasty or has a planned total knee arthroplasty in the next 24 months.

Trial design

Design model

Parallel

Treatments tested in this trial

  • Hip Innovation Technology Reverse Hip Replacement System

    Device

    Implantation of the Investigational Device is performed using standard surgical procedures for THA, as described in the Instructions for Use (IFU). The control hip systems will be implanted in accordance with their respective IFU, which are also in line with standard surgical approaches for THA.

  • Already-marketed semi-constrained uncemented hip systems using a metal-on-XLPE or ceramic-on-XLPE bearing combination.

    Device

    Control Arm subjects will receive one of the already-marketed semi-constrained uncemented hip systems using a metal-on-XLPE or ceramic-on-XLPE bearing combination. * DePuy Synthes CORAIL® or DePuy Synthes ACTIS™ Total Hip System in combination with DePuy Synthes PINNACLE™Complete Acetabular Hip System. * Zimmer Biomet Taperloc®, Zimmer Z1™ Cementless Femoral Hip Stem System or Zimmer Biomet AVENIR® Complete Hip System in combination with Zimmer Biomet G7® Acetabular System. * Stryker Accolade® II femoral stem in combination with Stryker Trident® II acetabular cup. * Smith \& Nephew ANTHOLOGY® Total Hip System in combination with Smith \& Nephew R3® Acetabular System

Treatment groups

288 Participants
are divided into 2 treatment groups
Group A: Experimental Arm - HIT Reverse HRSExperimental treatment 1 intervention
Group B: Control ArmActive comparator 1 intervention

Trial outcomes

Primary outcomes

1

OHS of 36 or more at the 2-year follow-up visit

The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).

Time frame
24 months
2

Change in OHS of 12 or more at the 2-year follow-up visit compared to baseline.

The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).

Time frame
24 months
3

Radiologic success at the 2-year follow-up

Defined as * No radiolucencies greater than 2 mm in 50% or more of the cup or stem zones; * No femoral or acetabular subsidence greater than or equal to 5 mm from baseline; * No acetabular cup inclination change greater than 4 degrees from baseline.

Time frame
24 months
4

Absence of revision at the 2-year follow-up visit

No reoperations that led to removal or replacement of any of the acetabular or femoral components.

Time frame
24 months

Secondary outcomes

1

Change in OHS compared to baseline at all follow-up visits.

The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).

Time frame
24 months
2

Change in HHS compared to baseline at all follow-up visits.

The Harris Hip Score (HHS) is a clinician-based 10-item questionnaire used to evaluate the results of hip surgery. It is comprised of 4 sub-sections yielding a maximum possible score of 100 points collectively, including pain (44 points), function (47 points), absence of deformity (4 points), and range of motion (5 points). Measured as Cohen's effect size, the improvements of THA are in the range of 3 and higher (Singh et al., 2016). The traditional categorization of HHS is: \< 70, poor; 70-79, fair; 80-89, good; and 90-100, excellent.

Time frame
24 months

Other outcomes

1

Exploratory Endpoints: OHS at all follow-up visits

The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).

Time frame
10 year
2

Exploratory Endpoints: HHS sub-scores (Pain, Function, Absence of Deformity, Range of Motion) at all follow-up visits.

The Harris Hip Score (HHS) is a clinician-based 10-item questionnaire used to evaluate the results of hip surgery. It is comprised of 4 sub-sections yielding a maximum possible score of 100 points collectively, including pain (44 points), function (47 points), absence of deformity (4 points), and range of motion (5 points). Measured as Cohen's effect size, the improvements of THA are in the range of 3 and higher (Singh et al., 2016). The traditional categorization of HHS is: \< 70, poor; 70-79, fair; 80-89, good; and 90-100, excellent.

Time frame
10 year
3

Exploratory Endpoints: Device survivorship

Device survivorship is measured as a cardinal event. Any reoperation in which the entire device or part of the device has been removed or replaced will be defined as a device revision.

Time frame
10 year
4

Exploratory Endpoints: Radiographic evaluation of linear penetration of the femoral head at the 2-year follow-up visit.

This study requires anteroposterior (AP)-pelvis and lateral (frog-leg or Lauenstein position) views obtained in the supine position on all subjects. An additional AP-hip image will be needed only if the AP-pelvis cannot provide a complete implant image. Radiographs will be obtained at the immediate postoperative follow-up (AP-pelvis only), and at the 6-week, 6-month, 1-year, 2-year, 3-year, 4-year, 5-year, and 10-year follow-up visits. Radiographs will be transferred to the Central Radiology Lab for processing and evaluation by an independent radiologist.

Time frame
10 year

Sponsors and contacts