About this trial
The purpose of this study is to determine if the use of the HIT Reverse Hip Replacement System is safe and effective in patients undergoing total hip replacement. We will determine whether it is safe and effective by comparing it to the control devices.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Patient requires primary THA due to NIDJD or any of its composite diagnoses such as osteoarthritis, avascular necrosis, posttraumatic arthritis, slipped capital epiphysis, fracture of the pelvis, and diastrophic variant requiring unilateral primary THA.
Patient is between 50 (inclusive) and 75 (inclusive) years of age at the time of enrollment.
Patient has preoperative medical clearance and is free from or treated for medical conditions that would pose excessive operative risk.
Patient has a signed and dated Informed Consent Form (ICF).
Disqualifiers
Patient has received earlier THA, hemi-arthroplasty, or fusion in the contralateral hip in the last 24 months.
Patient has received earlier THA, hemi-arthroplasty, or fusion in the contralateral hip more than 24 months ago and has a contralateral hip-specific pain rating of ≥2 on a Numeric Rating Scale 0-10.
Patient needs bilateral hip replacement or has a planned THA on the contralateral hip joint in the next 24 months.
Patient needs knee arthroplasty or has a planned total knee arthroplasty in the next 24 months.
Trial design
Parallel
Treatments tested in this trial
Hip Innovation Technology Reverse Hip Replacement System
DeviceImplantation of the Investigational Device is performed using standard surgical procedures for THA, as described in the Instructions for Use (IFU). The control hip systems will be implanted in accordance with their respective IFU, which are also in line with standard surgical approaches for THA.
Already-marketed semi-constrained uncemented hip systems using a metal-on-XLPE or ceramic-on-XLPE bearing combination.
DeviceControl Arm subjects will receive one of the already-marketed semi-constrained uncemented hip systems using a metal-on-XLPE or ceramic-on-XLPE bearing combination. * DePuy Synthes CORAIL® or DePuy Synthes ACTIS™ Total Hip System in combination with DePuy Synthes PINNACLE™Complete Acetabular Hip System. * Zimmer Biomet Taperloc®, Zimmer Z1™ Cementless Femoral Hip Stem System or Zimmer Biomet AVENIR® Complete Hip System in combination with Zimmer Biomet G7® Acetabular System. * Stryker Accolade® II femoral stem in combination with Stryker Trident® II acetabular cup. * Smith \& Nephew ANTHOLOGY® Total Hip System in combination with Smith \& Nephew R3® Acetabular System
Treatment groups
Trial outcomes
Primary outcomes
OHS of 36 or more at the 2-year follow-up visit
The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).
Change in OHS of 12 or more at the 2-year follow-up visit compared to baseline.
The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).
Radiologic success at the 2-year follow-up
Defined as * No radiolucencies greater than 2 mm in 50% or more of the cup or stem zones; * No femoral or acetabular subsidence greater than or equal to 5 mm from baseline; * No acetabular cup inclination change greater than 4 degrees from baseline.
Absence of revision at the 2-year follow-up visit
No reoperations that led to removal or replacement of any of the acetabular or femoral components.
Secondary outcomes
Change in OHS compared to baseline at all follow-up visits.
The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).
Change in HHS compared to baseline at all follow-up visits.
The Harris Hip Score (HHS) is a clinician-based 10-item questionnaire used to evaluate the results of hip surgery. It is comprised of 4 sub-sections yielding a maximum possible score of 100 points collectively, including pain (44 points), function (47 points), absence of deformity (4 points), and range of motion (5 points). Measured as Cohen's effect size, the improvements of THA are in the range of 3 and higher (Singh et al., 2016). The traditional categorization of HHS is: \< 70, poor; 70-79, fair; 80-89, good; and 90-100, excellent.
Other outcomes
Exploratory Endpoints: OHS at all follow-up visits
The Oxford Hip Score (OHS) is a 12-item patient-reported outcome measure designed and developed to assess function and pain in patients undergoing hip replacement surgery. It ranges from 0 (poor) to 48 (best).
Exploratory Endpoints: HHS sub-scores (Pain, Function, Absence of Deformity, Range of Motion) at all follow-up visits.
The Harris Hip Score (HHS) is a clinician-based 10-item questionnaire used to evaluate the results of hip surgery. It is comprised of 4 sub-sections yielding a maximum possible score of 100 points collectively, including pain (44 points), function (47 points), absence of deformity (4 points), and range of motion (5 points). Measured as Cohen's effect size, the improvements of THA are in the range of 3 and higher (Singh et al., 2016). The traditional categorization of HHS is: \< 70, poor; 70-79, fair; 80-89, good; and 90-100, excellent.
Exploratory Endpoints: Device survivorship
Device survivorship is measured as a cardinal event. Any reoperation in which the entire device or part of the device has been removed or replaced will be defined as a device revision.
Exploratory Endpoints: Radiographic evaluation of linear penetration of the femoral head at the 2-year follow-up visit.
This study requires anteroposterior (AP)-pelvis and lateral (frog-leg or Lauenstein position) views obtained in the supine position on all subjects. An additional AP-hip image will be needed only if the AP-pelvis cannot provide a complete implant image. Radiographs will be obtained at the immediate postoperative follow-up (AP-pelvis only), and at the 6-week, 6-month, 1-year, 2-year, 3-year, 4-year, 5-year, and 10-year follow-up visits. Radiographs will be transferred to the Central Radiology Lab for processing and evaluation by an independent radiologist.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.