Slew Rate As a Predictor for Optimal Lead Fixation

Trial statusRecruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
Age18-100
SponsorPaolo Pastori, MD

About this trial

The goal of this observational study is to identify the most reliable electrical parameter measured at the time of implantation that can predict optimal lead fixation and long-term lead performance in patients undergoing transvenous pacemaker or implantable cardioverter-defibrillator (ICD) implant.

The main questions it aims to answer are:

* Is there a correlation between the slew rate measured at the time of lead implantation and the sensing measured at the 12-month follow-up? * Is there a correlation between all electrical parameters recorded at implantation (including slew rate, current of injury, acute sensing, acute impedance, and acute pacing threshold) and the electrical parameters assessed at 12 months post-implantation (specifically chronic sensing, chronic impedance, and chronic pacing threshold)? Participants will undergo regular follow-up evaluations, with device interrogation visits scheduled at 1 and 12 months post-implantation as part of routine clinical care.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Implantation of a transvenous pacemaker with active lead fixation.

Implantation of a transvenous implantable cardioverter-defibrillator (ICD) with active lead fixation.

Disqualifiers

Pregnancy.

Age under 18 years.

Implantation of a subcutaneous ICD (S-ICD).

Implantation of a leadless pacemaker.

Trial population

Patients admitted to the hospital with an indication for scheduled or urgent transvenous pacemaker or implantable cardioverter-defibrillator (ICD) implantation with active lead fixation

Trial design

Design model

Cohort

Time perspective

Prospective

Treatments tested in this trial

Not listed

Trial groups

516 Participants
are grouped into 1 trial group
Group A: All patients

Trial outcomes

Primary outcomes

1

Slew rate value of active fixation leads (pacing or defibrillator leads) measured at the time of implantation

The slew rate value of active fixation leads (pacing or defibrillator leads) measured at the time of implantation will be automatically measured in millivolts (mV) using a Medtronic pacing system analyzer (PSA) (Model 2290) prior to helix deployment for active lead fixation. Both pacemaker and implantable cardioverter-defibrillator (ICD) leads will be included in the study. The lead fixation site (right ventricular apex or septum) and the implantation approach (traditional pacing or conduction system pacing) will be determined at the physician's discretion.

Time frame
After the enrollment, at the time of implantation, prior to helix deployment for active lead fixation

Secondary outcomes

1

Current of injury of active fixation leads (pacing or defibrillator leads) measured at the time of implantation

Characteristics of the current of injury-including total intracardiac EGM duration (msec), maximum ST elevation (mm), and maximum ST elevation 80 msec after the first EGM deflection (mm)-will be manually measured using the Medtronic PSA Model 2290 immediately after helix deployment during implantation

Time frame
After the enrollment, at the time of implantation, immediately following helix deployment for active lead fixation
2

Sensing of active fixation leads (pacing or defibrillator leads)

Sensing (mV) will be automatically measured using the PSA 2290 before and after helix deployment, as well as after lead fixation, followed by assessment via the device programmer at the end of the implant procedure and prior to discharge. During scheduled device interrogation visits at 1 and 12 months post-implant, sensing will be evaluated automatically using the device programmer.

Time frame
From enrollment at the time of implantation to the end of follow-up at 12 months
3

Capture threshold of active fixation leads (pacing or defibrillator leads)

Capture threshold (volts with a fixed impulse duration of 0.4 msec) will be manually measured using the PSA 2290 before and after helix deployment, as well as after lead fixation. This will be followed by assessment via the device programmer at the end of the implant procedure and prior to discharge. During scheduled device interrogation visits at 1 and 12 months post-implant, capture threshold will be manually evaluated using the device programmer.

Time frame
From enrollment at the time of implantation to the end of follow-up at 12 months
4

Impedance of active fixation leads (pacing or defibrillator leads)

Impedance (ohm) will be automatically measured using the PSA 2290 before and after helix deployment, as well as after lead fixation, followed by assessment via the device programmer at the end of the implant procedure and prior to discharge. During scheduled device interrogation visits at 1 and 12 months post-implant, impedance will be evaluated automatically using the device programmer.

Time frame
From enrollment at the time of implantation to the end of follow-up at 12 months

Other outcomes

1

Lead dislodgement

Any lead dislodgement (0 = No; 1 = Yes) occurring within the first 12 months post-implantation will be documented

Time frame
From the implantation to the end of follow-up at 12 months
2

Significant variation in electrical parameters

Any significant variation in electrical parameters occurring within the first 12 months post-implantation during follow-up will be documented, defined as follows: Pacing threshold: increase of \>1.5 V (with a fixed impulse duration of 0.4 msec) Sensing: decrease to \<4 mV Impedance: increase to \>1500 ohms

Time frame
From the implantation to the end of follow-up at 12 months

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Paolo Pastori, MD

Lead sponsor

Azienda Unità Sanitaria Locale di Parma

Sponsor institution

University Hospital of Ferrara

Collaborator

Azienda USL Reggio Emilia - IRCCS

Collaborator

Azienda Unita Sanitaria Locale di Piacenza

Collaborator