Smartphone Camera Measurement of Breathing-Related Heart Rate Variability Compared With ECG in Healthy Adults

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age19-39
SponsorBELL Therapeutics Inc.

About this trial

This study examines how accurately a smartphone can measure the rhythmic rise and fall of heart rate that occurs with breathing (respiratory heart rate variability).

Healthy adult volunteers attend one laboratory visit of about 90 minutes in Seoul, Republic of Korea. While seated, each participant breathes freely for 5 minutes, then follows audio-guided breathing at seven slow rates between 4 and 7 breaths per minute in random order and finally at 15 breaths per minute, with 2-minute rests between conditions (31 minutes of recording and 16 minutes of rest). During every condition, a single iPhone measures the face with its front camera (contact-free measurement) and a fingertip lightly placed over its rear camera lens (contact measurement) at the same time; no video is stored. A chest ECG and a respiratory belt record continuously throughout the session as references.

The main question is how closely the breathing-related heart rate variability amplitude measured from the face agrees with the ECG during slow paced breathing. The study also compares the fingertip measurement with the ECG, reports how often the smartphone measurements fail to produce a usable value, evaluates standard heart rate variability measures, and explores whether shorter recordings would give similar agreement and how breathing rate affects agreement.

The smartphone app is a research prototype used only for data collection. This study does not test a treatment, and its measurements are not used to diagnose or treat participants.

Eligibility criteria

This trial accepts healthy volunteers

Qualifiers

Adults aged 19 to 39 years

Able to understand the study explanation and give written informed consent

Able to remain seated for about 90 minutes and to follow audio-guided breathing at rates as slow as 4 breaths per minute

Disqualifiers

Diagnosed cardiac arrhythmia or conduction disorder (for example, atrial fibrillation)

Implanted cardiac pacemaker or implantable cardioverter-defibrillator

Current cardiopulmonary symptoms (for example, chest pain, shortness of breath or palpitations)

Any condition that, in the investigator's judgment, makes slow paced breathing unsafe or uncomfortable

Trial design

Design model

Single group

Treatments tested in this trial

  • Smartphone simultaneous facial and fingertip photoplethysmography (research app)

    Device

    Research-prototype iOS data-collection app on a single iPhone 17 Pro. The front camera measures the face for contact-free remote photoplethysmography; the rear main camera, with the flashlight on, measures a fingertip lightly placed over the lens for contact photoplethysmography. No video is stored; the app tracks both cheek regions and saves region-averaged signals with frame timestamps for later analysis. The app plays the breathing guide. It is not used for diagnosis, and readings shown during recording are not used as study results.

  • Audio-guided paced breathing

    Behavioral

    Fixed-rate audio cues at 15.0 and at 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 and 7.0 breaths/min, with inhalation taking 40% and exhalation 60% of each cycle, at a comfortable depth without breath holding. A free-breathing condition without cues is included. Participants may rest or stop at any time.

Treatment groups

36 Participants
are divided into 1 treatment group
Group A: Smartphone and Reference Measurement During Free and Paced BreathingExperimental treatment 2 interventions

Trial outcomes

Primary outcomes

1

Log Ratio of Facial Remote Photoplethysmography to ECG Respiratory Heart Rate Variability Amplitude During Slow Paced Breathing

ln(A\_pp face / A\_pp ECG) for each slow paced-breathing condition (4.0, 4.5, 5.0, 5.5, 6.0, 6.5 and 7.0 breaths/min). A\_pp is the peak-to-peak amplitude (ms) of the first harmonic of inter-beat intervals at the guided breathing frequency, estimated separately for the face and the ECG within the same window (60-180 s after the start of each 3-minute condition and 60-300 s for the 5-minute 6.0 breaths/min condition). Unit: natural-log ratio (unitless; 0 means no difference); +0.1 means a facial amplitude about 10% larger than the ECG amplitude. The reported value is the bias, the mean log ratio pooled across participants and the seven rates, with repeated-measures 95% limits of agreement and participant-level confidence intervals. Conditions without a valid facial estimate are counted in the facial measurement failure outcome.

Time frame
Seven slow paced-breathing conditions (3 minutes each; 6.0 breaths/min 5 minutes) during a single study visit (Day 1)

Secondary outcomes

1

Log Ratio of Fingertip Contact Photoplethysmography to ECG Respiratory Heart Rate Variability Amplitude During Slow Paced Breathing

ln(A\_pp fingertip / A\_pp ECG), with A\_pp, analysis windows and summary defined as in the primary outcome: the bias pooled across participants and the seven slow rates, with repeated-measures 95% limits of agreement and participant-level confidence intervals. Unit: natural-log ratio (unitless; 0 means no difference). Conditions without a valid fingertip estimate are counted in the fingertip measurement failure outcome.

Time frame
Seven slow paced-breathing conditions during a single study visit (Day 1)
2

Within-Participant Difference Between Facial-to-ECG and Fingertip-to-ECG Absolute Log Ratios of Respiratory Heart Rate Variability Amplitude During Slow Paced Breathing

For each participant, \|ln(A\_pp face / A\_pp ECG)\| minus \|ln(A\_pp fingertip / A\_pp ECG)\| in the same window, averaged over the slow paced-breathing conditions in which both sensors give a valid estimate. Unit: natural-log ratio (unitless); positive values mean a larger facial error. The reported value is the mean across participants with a participant-level 95% confidence interval. Each sensor is also compared with the ECG separately, because agreement between face and fingertip does not imply agreement with the ECG.

Time frame
Seven slow paced-breathing conditions during a single study visit (Day 1)
3

Percentage of Attempted Conditions Without a Valid Facial Estimate

Percentage of all originally attempted breathing conditions, pooled across participants, in which the facial pathway yields no valid estimate under the quality rules fixed before evaluation data are analyzed. The denominator is every attempted condition, including conditions whose ECG is unsuitable as a reference. Failed conditions are not removed, and the reason for each failure is reported.

Time frame
All nine conditions during a single study visit (Day 1)
4

Percentage of Attempted Conditions Without a Valid Fingertip Estimate

Percentage of all originally attempted breathing conditions, pooled across participants, in which the fingertip pathway yields no valid estimate under the quality rules fixed before evaluation data are analyzed. The denominator is every attempted condition, including conditions whose ECG is unsuitable as a reference. Failed conditions are not removed, and the reason for each failure is reported.

Time frame
All nine conditions during a single study visit (Day 1)

Other outcomes

1

Percentage of Participants Whose Facial and ECG Estimates Give the Same Breathing Rate With Maximal Respiratory Heart Rate Variability Amplitude (Exploratory)

For each participant, the rate between 4.0 and 7.0 breaths/min with the largest A\_pp is identified separately from the face and from the ECG. The outcome is the percentage of participants for whom the two rates are the same, among participants with valid estimates at all seven rates from both sources. A maximum at 4.0 or 7.0 breaths/min is flagged as a grid boundary. This rate is not treated as a physiological resonance frequency or an optimal treatment rate.

Time frame
Seven slow paced-breathing conditions during a single study visit (Day 1)
2

Percentage of Participants Whose Fingertip and ECG Estimates Give the Same Breathing Rate With Maximal Respiratory Heart Rate Variability Amplitude (Exploratory)

For each participant, the rate between 4.0 and 7.0 breaths/min with the largest A\_pp is identified separately from the fingertip and from the ECG. The outcome is the percentage of participants for whom the two rates are the same, among participants with valid estimates at all seven rates from both sources. A maximum at 4.0 or 7.0 breaths/min is flagged as a grid boundary. This rate is not treated as a physiological resonance frequency or an optimal treatment rate.

Time frame
Seven slow paced-breathing conditions during a single study visit (Day 1)
3

Absolute Difference Between Belt-Measured and Target Breathing Rate During Paced Breathing

Absolute difference (breaths/min) between the breathing rate measured by the respiratory belt and the target rate, averaged over the eight paced conditions for each participant and summarized as the mean across participants. Values for each target rate are supplementary.

Time frame
All eight paced conditions during a single study visit (Day 1)
4

Number of Participants With Symptoms Requiring a Pause or Stop

Symptoms such as dizziness, tingling, chest discomfort, breathlessness, or warmth or pain at the fingertip that lead to pausing or stopping a recording.

Time frame
During the single study visit (Day 1)